A new method for measuring aerosol nebulizer output using radioactive tracers.

نویسندگان

  • A A Gatnash
  • S T Chandler
  • C K Connolly
چکیده

Reproducibility and comparability of bronchial challenge tests depends critically on accurate assessment of nebulizer output. Evaporation during nebulization means that simple weighing is inaccurate, overestimating the delivered dose of active ingredient. We wanted to quantify this effect in the context of intermittent nebulization, using a dosimeter as used in bronchial provocation tests. Output of three types of nebulizer, from the MEFAR dosimeter, was measured by radioactive tracer, using a standard solution of technetium-99m-pertechnetate (1.5 kBq x mL(-1)) in 4 mL of normal saline. The aerosol was impacted by suction onto a microfilter, and the radioactivity measured. Nebulizers were weighed before and after nebulization. Ratio of nebulized volume calculated from the radioactivity on the filter, to the total volume loss by weight, was expressed as nebulized ratio. The effect on output of two concentrations of methacholine, two tracers of different weights, and change in temperature, were assessed. Nebulized ratio varied between 44.1-71.6%. Results were more consistent within the same type of nebulizer than between different makes. Neither changes in molar concentration nor molecular weight affected nebulizer output or nebulized ratio. Mean nebulized ratio was 58.5%, showing that calibration by weighing, overestimates the delivered dose by a factor of approximately two. Measuring radioactivity eluted from a microfilter, onto which nebulized output had been impacted proved to be a satisfactory method of calibration.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Measuring nebulizer output. Aerosol production vs gravimetric analysis.

STUDY OBJECTIVES The function of jet nebulizers has been measured traditionally by gravimetric methods, i.e., by weighing nebulizers before and after nebulization. Newer techniques measure aerosol output directly by analyzing aerosolized drug or tracer, i.e., radioactive 99mTc. Because of evaporation, the equivalence of these methods is uncertain. The aim of this study was to determine if the g...

متن کامل

Nebulizer calibration using lithium chloride: an accurate, reproducible and user-friendly method.

Conventional gravimetric (weight loss) calibration of jet nebulizers overestimates their aerosol output by up to 80% due to unaccounted evaporative loss. We examined two methods of measuring true aerosol output from jet nebulizers. A new adaptation of a widely available clinical assay for lithium (determined by flame photometry, LiCl method) was compared to an existing electrochemical method ba...

متن کامل

A device to validate concentration measured by direct reading instruments for aerosols

ii To: My parents Khalid and Sarah for their never ending support. My wife Alaa for her encouragement, patience, and believe in me. My children Khaled, Ahmad, and Sarah for bringing joy to my life. iii ACKNOWLEDGMENTS I would like to thank my advisor Dr. Thomas Peters for his support and guidance. I would also like to thank my committee members Dr. Patrick O'Shaughnessy and Dr. Renée Anthony fo...

متن کامل

Unattached radon progeny as an experimental tool for dosimetry of nanoaerosols: proposed method and research strategy.

In this paper the authors discuss a method using 1-nm particulate radon decay products as an experimental tool in the study of local lung deposition and dosimetry for nanoaerosols. The study of aerosol exposure and dosimetry measurements, and related quantitative assessment of health effects, are important to the understanding of the consequences of air pollution, and are discussed widely in th...

متن کامل

Variation in nebulizer output and weight output from Mefar dosimeter: implications for multicentre studies.

We wish to comment upon the paper of CHINN et al. [1] in which both weight and aerosol output data were given for 193 Mefar nebulizers from 27 centres involved in the European Community Respiratory Health Survey (ECRHS). A single centre in Turin determined nebulizer output using the sodium fluoride (NaF) tracer method described by DENNIS et al. [2]. The weight output was given as 11 mg·s-1 with...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 12 2  شماره 

صفحات  -

تاریخ انتشار 1998