Dosimetric characteristics of electron beams produced by two mobile accelerators, Novac7 and Liac, for intraoperative radiation therapy through Monte Carlo simulation

نویسندگان

  • Sergio Righi
  • Evis Karaj
  • Giuseppe Felici
  • Fabio Di Martino
چکیده

The Novac7 and Liac are linear accelerators (linacs) dedicated to intraoperative radiation therapy (IORT), which produce high energy, very high dose-per-pulse electron beams. The characteristics of the accelerators heads of the Novac7 and Liac are different compared to conventional electron accelerators. The aim of this work was to investigate the specific characteristics of the Novac7 and Liac electron beams using the Monte Carlo method. The Monte Carlo code BEAMnrc has been employed to model the head and simulate the electron beams. The Monte Carlo simulation was preliminarily validated by comparing the simulated dose distributions with those measured by means of EBT radiochromic film. Then, the energy spectra, mean energy profiles, fluence profiles, photon contamination, and angular distributions were obtained from the Monte Carlo simulation. The Spencer-Attix water-to-air mass restricted collision stopping power ratios (sw,air) were also calculated. Moreover, the modifications of the percentage depth dose in water (backscatter effect) due to the presence of an attenuator plate composed of a sandwich of a 2 mm aluminum foil and a 4 mm lead foil, commonly used for breast treatments, were evaluated. The calculated sw,air values are in agreement with those tabulated in the IAEA TRS-398 dosimetric code of practice within 0.2% and 0.4% at zref (reference depth in water) for the Novac7 and Liac, respectively. These differences are negligible for practical dosimetry. The attenuator plate is sufficient to completely absorb the electron beam for each energy of the Novac7 and Liac; moreover, the shape of the dose distribution in water strongly changes with the introduction of the attenuator plate. This variation depends on the energy of the beam, and it can give rise to an increase in the maximum dose in the range of 3%-9%.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator

Background: One of the main problems of dedicated IORT accelerators is to determine dosimetric characteristics of the electron beams. Monte Carlo simulation of IORT accelerator head and produced beam will be useful to improve the accuracy of beam dosimetry.Materials and Methods: Liac accelerator head was modeled using the BEAMnrcMonte Carlo simulation system. Phase-space files were generated at...

متن کامل

اعتبارسنجی سر شتاب‌دهندۀ پرتودرمانی حین عمل‌ جراحی LIAC با استفاده از روش مونت‌کارلو

With the development of science and technology in the medical field and need more than ever to accuracy and quality of treatment has been causer increase different radiotherapy methods. Almost the past two decade’s intraoperative radiation therapy (IORT) as one of the new techniques used to treat cancer patients. One of the problems with this method is obtaining accurate dosimetry, becaus...

متن کامل

Physical characteristics of electron beam from conventional and beam shaper IOERT applicator: A comparison study

Introduction: Intraoperative electron radiation therapy (IOERT) is one of the cancer treatment techniques that delivers high doses to tumor bed during surgery. IOERT can be performed by either conventional LINACs or dedicated IORT accelerators such as LIAC (Light Intraoperative Accelerator). Two types of applicators can be used with LIAC dedicated accelerator including conventi...

متن کامل

شبیه‌سازی شتاب‌ دهنده پرتو درمانی حین عمل LIAC به همراه اپلیکاتور شکل دهنده باریکه بااستفاده از محاسبات مونت‌کارلو

Intraoperative electron radiotherapy is one of the radiotherapy methods that delivers a high single fraction of radiation dose to the patient in one session during the surgery. Beam shaper applicator is one of the applicators that is recently employed with this radiotherapy method. This applicator has a considerable application in treatment of large tumors. In this study, the dosimetric charact...

متن کامل

Nuclear Physics Comparison of Measured and Monte Carlo Calculated Dose Distributions for the Novac7® Linear Accelerator

Percentage depth doses (PDDs) and transverse dose profiles (TDPs) were calculated by a Monte Carlo simulation of electron beams of different energies (3, 5, 7 and 9 MeV) produced by a NOVAC7® accelerator. The NOVAC7® is a linear accelerator designed for Intraoperative Radiation Therapy (IORT). The PDDs and TDPs were also experimentally determined. Both calculations and measurements were done on...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 14  شماره 

صفحات  -

تاریخ انتشار 2013