Sensitivity and daily quality control of a mobile PET/CT scanner operating in 3-dimensional mode.
نویسندگان
چکیده
UNLABELLED This study investigated the stability of the sensitivity of a mobile PET/CT scanner and tested a phantom experiment to improve on the daily quality control recommendations of the manufacturer. Unlike in-house scanners, mobile PET/CT devices are subjected to a harsher, continuously changing environment that can alter their performance. The parameter of sensitivity was investigated because it reflects directly on standardized uptake value, a key factor in cancer evaluation. METHODS A (68)Ge phantom of known activity concentration was scanned 6 times a month for 11 consecutive months using a mobile PET/CT scanner that operates in 3-dimensional mode only. The scans were acquired as 2 contiguous bed positions, with raw data obtained and reconstructed using parameters identical to those used for oncology patients, including CT-extracted attenuation coefficients and decay, scatter, geometry, and randoms corrections. After visual inspection of all reconstructed images, identical regions of interest were drawn on each image to obtain the activity concentration of individual slices. The original activity concentration was then decay-corrected to the scanning day, and the percentage sensitivity of the slice was calculated and graphed. The daily average sensitivity of the scanner, over 11 consecutive months, was also obtained and used to evaluate the stability of sensitivity. RESULTS Our particular scanner showed a daily average sensitivity ranging from -8.6% to 6.5% except for one instance, when the sensitivity dropped by an unacceptable degree, 34.8%. CONCLUSION Our 11-mo follow-up of a mobile PET/CT scanner demonstrated that its sensitivity remained within acceptable clinical limits except for one instance, when the scanner had to be serviced before patients could be imaged. To enhance our confidence in the uniformity of sensitivity across slices, we added a phantom scan to the daily quality control recommendations of the manufacturer.
منابع مشابه
An overview of clinical PET/CT
This article is intended to provide an overview of various aspects of clinical PET/CT. These include discussions of: (i) Important areas of clinical application; (ii) Opportunities in clinical research; (iii) Scanner and operating-mode considerations (e.g. BGO vs. LSO, LYSO or GSO scanners, 2D vs. 3D imaging). (iv) Study-specific considerations (e.g. patient preparation and positioning is...
متن کاملPerformance comparison of four commercial GE discovery PET/CT scanners: A monte carlo study using GATE
Combined PET/CT scanners now play a major role in medicine for in vivo imaging in oncology, cardiology, neurology, and psychiatry. As the performance of a scanner depends not only on the scintillating material but also on the scanner design, with regards to the advent of newer scanners, there is a need to optimize acquisition protocols as well as to compare scanner ...
متن کاملImpact of Time-of-Flight PET/CT with a Large Axial Field of View for Reducing Whole-Body Acquisition Time.
UNLABELLED The aim of this study was to evaluate the imaging performance of 39- and 52-ring time-of-flight (TOF) PET/CT scanners. We also assessed the potential of reducing the scanning time using a 52-ring TOF PET/CT scanner. METHODS PET/CT scanners with 39- and 52-ring lutetium oxyorthosilicate detectors were evaluated. The axial fields of view were 16.2 and 21.6 cm, respectively. We used a...
متن کاملDesign and performance evaluation of spheroid geometry for brain PET scanner using Monte Carlo modeling
Introduction: There has been a curiosity about the spheroid geometry for PET scanners developments since several years ago, therefore in this study, we are aiming to evaluate the performance of this geometry and compare its performance with cylindrical geometry using Monte Carlo simulation. Methods: We simulated a spheroid geometry with a radius of 199 mm...
متن کاملDetection of prostate cancer by an FDG-PET cancer screening program: results from a Japanese nationwide survey
Objective(s): The aim of this study was to analyze detection rates and effectiveness of 18F-fluorodeoxyglucose positron emission tomography (FDG-PET) cancer screening program for prostate cancer in Japan, which is defined as a cancer-screening program for subjects without known cancer. It contains FDG-PET aimed at detection of cancer at an early stage with or without additional screening tests ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of nuclear medicine technology
دوره 35 4 شماره
صفحات -
تاریخ انتشار 2007