An improvement in IMRT QA results and beam matching in linacs using statistical process control

نویسندگان

  • Justin D. Gagneur
  • Gary A. Ezzell
چکیده

The purpose of this study is to apply the principles of statistical process control (SPC) in the context of patient specific intensity-modulated radiation therapy (IMRT) QA to set clinic-specific action limits and evaluate the impact of changes to the multileaf collimator (MLC) calibrations on IMRT QA results. Ten months of IMRT QA data with 247 patient QAs collected on three beam-matched linacs were retrospectively analyzed with a focus on the gamma pass rate (GPR) and the average ratio between the measured and planned doses. Initial control charts and action limits were calculated. Based on this data, changes were made to the leaf gap parameter for the MLCs to improve the consistency between linacs. This leaf gap parameter is tested monthly using a MLC sweep test. A follow-up dataset with 424 unique QAs were used to evaluate the impact of the leaf gap parameter change. The initial data average GPR was 98.6% with an SPC action limit of 93.7%. The average ratio of doses was 1.003, with an upper action limit of 1.017 and a lower action limit of 0.989. The sweep test results for the linacs were -1.8%, 0%, and +1.2% from nominal. After the adjustment of the leaf gap parameter, all sweep test results were within 0.4% of nominal. Subsequently, the average GPR was 99.4% with an SPC action limit of 97.3%. The average ratio of doses was 0.997 with an upper action limit of 1.011 and a lower action limit of 0.981. Applying the principles of SPC to IMRT QA allowed small differences between closely matched linacs to be identified and reduced. Ongoing analysis will monitor the process and be used to refine the clinical action limits for IMRT QA.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The step‐and‐shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage

The goal of this work is to evaluate the dosimetric impact of an overshooting phenomenon in step-and-shoot IMRT delivery, and to demonstrate a novel method to mitigate the issue. Five pelvis IMRT patients treated on Varian 2100C EX linacs with larger than +4.5% phantom ion chamber point-dose difference relative to planned dose were investigated. For each patient plan, 5 fractions were delivered...

متن کامل

Statistical process control analysis for patient-specific IMRT and VMAT QA

This work applied statistical process control to establish the control limits of the % gamma pass of patient-specific intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) quality assurance (QA), and to evaluate the efficiency of the QA process by using the process capability index (Cpml). A total of 278 IMRT QA plans in nasopharyngeal carcinoma were measured with ...

متن کامل

A management method for the statistical results of patient-specific quality assurance for intensity-modulated radiation therapy

There are many reports concerning patient-specific quality assurance (QA) for intensity-modulated radiation therapy (IMRT). However, reports about the statistical results of QA are lacking. Management methods for the results of the QA are needed, even though we have the ESTRO group recommendation that a tolerance limit of 1.96 standard deviation (SD) be established in each institution. The purp...

متن کامل

A Feasibility Study of IMRT of Lung Cancer Using Gafchromic EBT3 Film

Background: Intensity modulated radiation therapy (IMRT) is an advanced method for delivery of three dimensional therapies, which provides optimal dose distribution with giving multiple nonuniform fluency to the patient. The complex dose distribution of IMRT should be checked to ensure that the accurate dose is delivered. Today, film dosimetry is a powerful tool for radiotherapy treatment Quali...

متن کامل

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 15  شماره 

صفحات  -

تاریخ انتشار 2014