Double-tilt in situ TEM holder with multiple electrical contacts and its application in MEMS-based mechanical testing of nanomaterials.

نویسندگان

  • Rodrigo A Bernal
  • Rajaprakash Ramachandramoorthy
  • Horacio D Espinosa
چکیده

MEMS and other lab-on-a-chip systems are emerging as attractive alternatives to carry out experiments in situ the electron microscope. However, several electrical connections are usually required for operating these setups. Such connectivity is challenging inside the limited space of the TEM side-entry holder. Here, we design, implement and demonstrate a double-tilt TEM holder with capabilities for up to 9 electrical connections, operating in a high-resolution TEM. We describe the operating principle of the tilting and connection mechanisms and the physical implementation of the holder. To demonstrate the holder capabilities, we calibrate the tilting action, which has limits of ±15°, and establish the insulation resistance of the electronics to be 36GΩ, appropriate for measurements of currents down to the nano-amp (nA) regime. Furthermore, we demonstrate tensile testing of silver nanowires using a previously developed MEMS device for mechanical testing, using the implemented holder as the platform for electronic operation and sensing. The implemented holder can potentially have broad application to other areas where MEMS or electrically-actuated setups are used to carry out in situ TEM experiments.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Quantitative in situ TEM tensile fatigue testing on nanocrystalline metallic ultrathin films.

A unique technique to perform quantitative in situ transmission electron microscopy (TEM) fatigue testing on ultrathin films and nanomaterials is demonstrated. The technique relies on a microelectromechanical system (MEMS) device to actuate a nanospecimen and measure its mechanical response. Compared to previously demonstrated MEMS-based in situ TEM techniques, the technique takes advantage of ...

متن کامل

Towards 3D crystal orientation reconstruction using automated crystal orientation mapping transmission electron microscopy (ACOM-TEM)

To relate the internal structure of a volume (crystallite and phase boundaries) to properties (electrical, magnetic, mechanical, thermal), a full 3D reconstruction in combination with in situ testing is desirable. In situ testing allows the crystallographic changes in a material to be followed by tracking and comparing the individual crystals and phases. Standard transmission electron microscop...

متن کامل

In situ transmission electron microscopy studies enabled by microelectromechanical system technology

We have designed and fabricated a standardized specimen holder that allows the operation of a microelectromechanical system (MEMS) device inside a transmission electron microscope (TEM). The details of the design and fabrication processes of the holder are presented. The sample loading mechanism is simple and allows reliable electrical contact to eight signal lines on the device. Using a MEMS-b...

متن کامل

Controlled In Situ Gas Reaction Studies of Catalysts at High Temperature and Pressure with Atomic Resolution

In situ reaction studies of catalyst materials using closed-cell environmental specimen holders have been shown to allow atomic resolution to be obtained on e.g. catalyst materials, at elevated temperatures and pressures [1-4]. These holders incorporate MEMS-based devices into the holder tip, which serve to encapsulate a gas layer between a thin film heater device and an ultra-thin amorphous Si...

متن کامل

A MEMS device for In Situ TEM/AFM/SEM/STM Testing of Carbon Nanotubes and Nanowires

A MEMS device for in-situ TEM/AFM/SEM/STM testing of nano structures is designed. Two deformation measurement methods are discussed: microscopic measurement and electrical measurement. For the first method, the device consists of a comb-drive actuator, calibration beam and specimen stage. The stiffness of the specimen can be well characterized by comparing the load-displacement relation before ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Ultramicroscopy

دوره 156  شماره 

صفحات  -

تاریخ انتشار 2015