Inappropriate elevation of bispectral index values in robot assisted thyroidectomy with electromyographic endotracheal tube -A case report-

نویسندگان

  • Mae-Hwa Kang
  • Chi-Bum In
  • Man-ho Kim
  • Kyoung-Ji Lim
  • Eun Young Park
  • Hyo Min Lee
  • Soo Kyung Lee
چکیده

We report a case of an erroneously elevated bispectral index (BIS) during robot assisted thyroidectomy using an electromyographic endotracheal tube (EMG tube), which is safe and useful for laryngeal electromyographic monitoring. Ten minutes after start of the operation, a sudden increase of BIS value up to 98 was noticed. The BIS values were not decreased to < 65 with supplemental anesthetics. The anesthetic method was changed from total intravenous anesthesia to balanced anesthesia. The BIS sensor and monitor were changed and other models were used. These interventions did not alter BIS values. BIS levels remained between 60 and 70 throughout the main procedure and intermittently increased to the mid-90s without any trace of poor signal quality. At the end of the surgery, the BIS values returned to normal range. The patient did not complain of intraoperative recall. Knowledge of potential interference from the use of an EMG tube must be considered when interpreting BIS.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Rupture of endotracheal tube cuff during robot-assisted endoscopic thyroidectomy -A case report-

We encountered a case of a rupture of an endotracheal tube cuff during robot-assisted thyroid surgery in a 35-year-old male patient. Two hours after commencing surgery, the bellows of the ventilator were not filled and a rupture of the endotracheal tube cuff was suspected. Once the robot-manipulator is engaged, the position of the operating table cannot be altered without removing it from the p...

متن کامل

Successful anesthetic management of a patient with thyroid carcinoma invading the trachea with tracheal obstruction, scheduled for total thyroidectomy.

We report a case of large thyroid carcinoma with tracheal and esophageal invasion who presented with preoperative stridor scheduled for total thyroidectomy and segmental tracheal resection. Careful and comprehensive preoperative anesthetic planning was done. Extracorporeal circulation membrane oxygenation (ECMO) was set up and running prior to induction under local anesthesia, due to an increas...

متن کامل

Ligation of the internal jugular vein increased regional cerebral oxygen saturation but decreased the bispectral index in a 72-year-old man: a case report

Regional cerebral oxygen saturation (rSO2) and the bispectral index (BIS) are used to detect cerebral perfusion abnormalities. However, whether rSO2 and BIS values change during ligation of the internal jugular vein (IJV) is unknown. We report a case in which BIS values were decreased, despite increased rSO2 during ligation of the IJV. A 72-year-old man was diagnosed with metastasis of renal ca...

متن کامل

Tongue necrosis: a rare complication of oral intubation.

OBJECTIVES (1) Present a unique case of partial necrosis of the dorsal tongue caused by an endotracheal tube; (2) highlight the importance of verifying proper endotracheal tube positioning during cases requiring prolonged intubation. METHODS Case report and literature review. RESULTS A 50 year-old man underwent total thyroidectomy and bilateral lymphadenectomies for papillary thyroid carcin...

متن کامل

Electromyographic activity falsely elevates the bispectral index.

THE bispectral index (BIS, Aspect Medical Systems, Framingham, MA) is a complex electroencephalographic (EEG) parameter that integrates several disparate descriptors of the EEG into a single variable and correlates behavioral assessments of sedation and hypnosis. We report two cases in which the bispectral index failed to measure depth of anesthesia but instead was a measure of electromyographi...