Accuracy and precision of the Axis-Shield Afinion hemoglobin A1c measurement device.

نویسندگان

  • Jamie R Wood
  • Brett M Kaminski
  • Craig Kollman
  • Roy W Beck
  • Callyn A Hall
  • Jason P Yun
  • Eda Cengiz
  • Michael J Haller
  • Krishna Hassan
  • Georgeanna J Klingensmith
  • William V Tamborlane
چکیده

BACKGROUND The Afinion HbA1c (Axis-Shield) is a newer point-of-care device for measurement of hemoglobin A1c (A1C) using a boronate affinity method unlike the more commonly used DCA immunoassay method (Siemens Medical Solutions Diagnostics). The Afinion's accuracy and precision, when compared with high-performance liquid chromatography (HPLC) and DCA methods, have not been established in pediatric practice settings. METHODS Capillary blood was collected from 700 subjects with diabetes mellitus at seven Pediatric Diabetes Consortium sites. Each subject's A1C was measured locally using Afinion and DCA devices, and by a central laboratory (University of Minnesota) using a Tosoh HPLC method. In addition, repeated measurements on six whole blood samples provided by the National Glycohemoglobin Standardization Program (NGSP) were taken at three clinical centers using the Afinion and DCA methods and centrally using the Tosoh HPLC method to assess the precision of each device. RESULTS The coefficient of variation for measurements of whole blood samples for precision analysis was 2% for Afinion, 3% for DCA, and 1% for HPLC. In the patient samples measured at the seven clinic sites, the Afinion generated higher A1C results than the HPLC (mean difference = +0.15; p < 0.001), while the DCA produced lower values (mean difference = -0.19; p < 0.001). The absolute differences with HPLC were similar for the Afinion and DCA (median 0.2%) with a slight advantage for the Afinion when compared with DCA (p < 0.001 by rank test). The DCA tended to read lower than HPLC, particularly at high A1C levels (p < 0.001), while the Afinion's accuracy did not vary by A1C. CONCLUSIONS When compared to the central laboratory HPLC method, the differences between the results of the Afinion and DCA devices are clinically insignificant, and the Afinion and DCA have similar accuracy and precision when used in pediatric practice settings.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Analysis of the accuracy and precision of the Axis-Shield Afinion hemoglobin A1c measurement device.

Point-of-care (POC) hemoglobin A1c measurement is now used by many physicians to make more timely decisions on therapy changes. A few studies have highlighted the drawbacks of some POC methods, e.g., poor precision and lot-to-lot variability. Evaluating performance in the clinical setting is difficult because there is minimal proficiency testing data on POC methods. In this issue of Journal of ...

متن کامل

Evaluation of the Afinion AS100 Point-of-Care Analyzer for Hemoglobin A1c

An evaluation of a new Afinion AS100 Analyzer was conducted to assess analytical performance. Precision was estimated by analyzing 2 control and 3 patient samples twice a day for 10 days. Accuracy was established by analysis of 6 samples from the National Glycated Hemoglobin Standardization Program for a 3-day period. Agreement was correlated to a laboratory method, the Variant II Turbo Hemoglo...

متن کامل

Evaluation of two HbA1c point-of-care analyzers.

BACKGROUND Measurement of HbA1c is the most important parameter to assess glycemic control in diabetic patients. Different point-of-care devices for HbA1c are available. The aim of this study was to evaluate two point-of-care testing (POCT) analyzers (DCA Vantage from Siemens and Afinion from Axis-Shield). We studied the bias and precision as well as interference from carbamylated hemoglobin. ...

متن کامل

Three of 7 hemoglobin A1c point-of-care instruments do not meet generally accepted analytical performance criteria.

BACKGROUND In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS Clinical and Laboratory Sta...

متن کامل

Selecting an A1C Point-of-Care Instrument

A1C point-of-care (POC) instruments benefit patients with diabetes by facilitating clinician decision making that results in significant glycemic improvements. Three National Glycohemoglobin Standardization Program (NGSP)-certified POC products are available in the United States: the handheld A1CNow (formerly manufactured by Bayer Diabetes Care but now made by Chek Diagnostics) and two bench-to...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of diabetes science and technology

دوره 6 2  شماره 

صفحات  -

تاریخ انتشار 2012