Dosimetric characterization of the M−15 high‐dose‐rate Iridium−192 brachytherapy source using the AAPM and ESTRO formalism
نویسندگان
چکیده
The Source Production & Equipment Co. (SPEC) model M-15 is a new Iridium-192 brachytherapy source model intended for use as a temporary high-dose-rate (HDR) brachytherapy source for the Nucletron microSelectron Classic afterloading system. The purpose of this study is to characterize this HDR source for clinical application by obtaining a complete set of Monte Carlo calculated dosimetric parameters for the M-15, as recommended by AAPM and ESTRO, for isotopes with average energies greater than 50 keV. This was accomplished by using the MCNP6 Monte Carlo code to simulate the resulting source dosimetry at various points within a pseudoinfinite water phantom. These dosimetric values next were converted into the AAPM and ESTRO dosimetry parameters and the respective statistical uncertainty in each parameter also calculated and presented. The M-15 source was modeled in an MCNP6 Monte Carlo environment using the physical source specifications provided by the manufacturer. Iridium-192 photons were uniformly generated inside the iridium core of the model M-15 with photon and secondary electron transport replicated using photoatomic cross-sectional tables supplied with MCNP6. Simulations were performed for both water and air/vacuum computer models with a total of 4 × 109 sources photon history for each simulation and the in-air photon spectrum filtered to remove low-energy photons belowδ = 10 keV. Dosimetric data, including D·(r,θ), gL(r), F(r,θ), φan(r), and φ-an, and their statistical uncertainty were calculated from the output of an MCNP model consisting of an M-15 source placed at the center of a spherical water phantom of 100 cm diameter. The air kerma strength in free space, SK, and dose rate constant, Λ, also was computed from a MCNP model with M-15 Iridium-192 source, was centered at the origin of an evacuated phantom in which a critical volume containing air at STP was added 100 cm from the source center. The reference dose rate, D·(r0,θ0) ≡ D· (1cm,π/2), is found to be 4.038 ± 0.064 cGy mCi-1 h-1. The air kerma strength, SK, is reported to be 3.632 ± 0.086 cGy cm2 mCi-1 g-1, and the dose rate constant, Λ, is calculated to be 1.112 ± 0.029 cGy h-1 U-1. The normalized dose rate, radial dose function, and anisotropy function with their uncertainties were computed and are represented in both tabular and graphical format in the report. A dosimetric study was performed of the new M-15 Iridium-192 HDR brachytherapy source using the MCNP6 radiation transport code. Dosimetric parameters, including the dose-rate constant, radial dose function, and anisotropy function, were calculated in accordance with the updated AAPM and ESTRO dosimetric parameters for brachytherapy sources of average energy greater than 50 keV. These data therefore may be applied toward the development of a treatment planning program and for clinical use of the source.
منابع مشابه
Dosimetric characterization of a high dose rate 192I source for brachytherapy application using Monte Carlo simulation and benchmarking with thermoluminescent dosimetry
Background: The purpose of this project was to derive the brachytherapy dosimetric functions described by American Association of Physicists in Medicine (AAPM) TG-43 U1 based on high dose rate 192I sources. Materials and Methods: The method utilized included both simulation of the designed Polymethyl methacrylate (PMMA) phantom using the Monte Carlo of MCNP4C and benchmarking of the simulation ...
متن کاملDetermination of Dosimetric characteristics of a New 192Ir-PDR Brachytherapy Source According to AAPM TG- 43 Protocol using Monte Carlo simulation technique
Introduction: 192Ir is one of the important sources frequently used in brachytherapy. Up to now, a lot of commercial models of this source have been made which Ir-192 has been recently added to them. The aim of the present study is to determine the dosimetric parameters of this new source model based on the recommendations of TG-43(U1) protocol using Monte Carlo simulation tech...
متن کاملTG-60 dosimetry parameters calculation for the β-emitter 153Sm brachytherapy source using MCNP
Background: The formalism recommended by Task Group 60 (TG-60) of the American Association of Physicists in Medicine (AAPM) is applicable for β sources. Radioactive biocompatible and biodegradable 153Sm glass seed without encapsulation is a β- emitter with a short half life and delivers a high dose rate to the tumor in the millimeter range. In this work the dosimetry parameters of the ...
متن کاملMonte Carlo calculation of the TG-43 dosimetric parameters of a new BEBIG Ir-192 HDR source.
BACKGROUND AND PURPOSE High dose rate (HDR) brachytherapy is a highly extended practice in clinical brachytherapy today. Quality dose rate distribution datasets of the HDR sources used in a clinical treatment are required. Because of the different source designs, a specific dosimetry dataset is required for each source model. In the recently published BRAPHYQS-ESTRO Report, an overview of avail...
متن کاملCalibration of iridium-192 source by ionization chamber for high dose rate brachytherapy
Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...
متن کامل