Comparison of the Radiation Dose from Cone Beam Computed Tomography and Multidetector Computed Tomography in Examinations of the Hand Vergleich der Strahlendosis von Cone-Beam Computertomografie und Multidetektor Computertomografie in Untersuchungen der Hand
نویسندگان
چکیده
Correspondence Dr. Jakob Neubauer Radiology, University Medical Center Freiburg Hugstetterstr. 55 79106 Freiburg Germany Tel.: ++ 49/7 61/3 8262 Fax: ++ 49/7 61/3 8300 [email protected] Abstract ▼ Purpose: Comparison of radiation dose of cone beam computed tomography (CBCT) and multidetector computed tomography (MDCT) in examinations of the hand. Materials and Methods: Dose calculations were carried out by means of Monte Carlo simulations in MDCT and CBCT. A corpse hand was examined in a 320-row MDCT scanner and a dedicated extremities CBCT scanner with standard protocols and multiple lowdose protocols. The image quality of the examinations was evaluated by 5 investigators using a Likert scale from 1 (very good) to 5 (very poor) regarding depiction of cortical bone, cancellous bone, joint surfaces, soft tissues and artifacts. For a sum of ratings of all structures < 50 a good overall image quality was expected. The studies with at least good overall image quality were comparedwith respect to the dose. Results: The dose of the standard examination was 13.21 (12.96 to 13.46 CI) mGy in MDCT and 7.15 (6.99 to 7.30 CI) mGy in CBCT. The lowest dose in a studywith good overall image quality was 4.54 (4.43 to 4.64 CI) mGy in MDCT and 5.72 (5.59 to 5.85 CI) mGy in CBCT. Conclusion: Although the dose of the standard protocols in the CBCT is lower than in the MDCT, the MDCT can realize a good overall image quality at a lower dose than the CBCT. Dose optimization of CT examination protocols for the hand is useful in both modalities, the MDCT has an even greater potential for optimization. Key points: ▶ Low dose examinations of the hand are feasible in CBCT and MDCT. ▶ In default settings CBCT has a lower dose than MDCT. ▶ MDCT enables a good image quality at a lower dose than CBCT. Citation Format: ▶ Neubauer J, Neubauer C, Gerstmair A et al. Comparison of the Radiation Dose from Cone Beam Computed Tomography and Multidetector Computed Tomography in Examinations of the Hand. Fortschr Röntgenstr 2016; 188: 488–493
منابع مشابه
High-Pitch Computed Tomography of the Lung in Pediatric Patients: An Intraindividual Comparison of Image Quality and Radiation Dose to Conventional 64-MDCT High-pitch Computertomografie der Lunge bei Kindern: Ein intraindividueller Vergleich der Bildqualität und Strahlendosis zur konventionellen 64-MDCT
Ziel: Ziel der Studie war die Häufigkeit von typischen Artefakten der high-pitch Computertomografie (HPCT) bei low-dose Untersuchungen von Kindern im Vergleich zur MDCT zu evaluieren und die effektive Dosis (Deff) zu berechnen. Material und Methoden: 35 Patienten (17 Jungen, 18 Mädchen; Durchschnittsalter 112±69 Monate) wurden in diese retrospektive Studie eingeschlossen. Bei allen Patienten wu...
متن کاملArtifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality
Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...
متن کاملThe Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography
Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...
متن کاملDosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography
Background: While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free.Objective: The aim of this study was to evaluate the exposure doses received by patients during maxillofacial imaging with CBCT.Methods: Entrance surface dose (ESD) was measured by using thermoluminescent dosimeters (TLDs) attached to the eyes...
متن کاملمقایسه دقت توموگرافی سهبعدی با اشعه مخروطی با توموگرافی کامپیوتری مرسوم در تشخیص بیماریهای سینوس پارانازال
Background: Paranasal sinus disease is one of the most common problems of patients that refer to ENT centers. Adding to clinical examination imaging plays an important role in diagnosis and treatment. Also the imaging of paranasal sinuses is necessary before surgery. Although computed tomography is the modality of choice for these areas, it has some disadvantages which the most important one is...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2016