Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT).
نویسندگان
چکیده
BACKGROUND Two electronic devices for self-measurement of blood pressure - a brachial monitor, the Omron M6, and a wrist monitor, the Omron R7 - were evaluated in two separate studies according to the International Protocol of the European Society of Hypertension. DESIGN The International Validation Protocol is divided into two phases: the first phase is performed on 15 selected participants (45 pairs of blood pressure measurements); if the device passes this phase, 18 supplementary participants are included (54 pairs of blood pressure measurements) making a total number of 33 participants (99 pairs of blood pressure measurements) on whom the final validation is performed. METHODS The same methodology recommended by the European Society of Hypertension protocol was applied for both studies. In each study and for each participant, four blood pressure measurements were taken simultaneously by two trained observers using mercury sphygmomanometers alternately with three measurements taken by the tested device. The difference between the blood pressure value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 pairs of blood pressure differences were classified into three categories (<or=5, <or=10 and <or=15 mmHg). The number of differences in each category was compared with the number required by the International Protocol. An individual analysis was then done to determine the number of comparisons <or=5 mmHg for each participant. At least 22 of the 33 participants should have two of their three comparisons <or=5 mmHg. RESULTS In both studies, the two tested devices passed the first and the second phases of the validation process. The average differences between the device and mercury sphygmomanometer readings were 0.8+/-2.7 and -1.9+/-3.3 mmHg for systolic and diastolic blood pressure, respectively, for the Omron M6 device, and 0.2+/-4.2 and 0.2+/-2.9 mmHg for systolic and diastolic blood pressure, respectively, for the Omron R7 device. For both devices, readings differing by less than 5, 10 and 15 mmHg for systolic and diastolic blood pressure values fulfill the recommendation criteria of the International Protocol as well as the individual analysis. CONCLUSIONS The Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT) devices fulfilled the validation recommendations of the International Protocol.
منابع مشابه
Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European Society of Hypertension International Protocol
BACKGROUND Four oscillometric devices, including the Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670, designed for self-blood pressure measurement (SBPM) were evaluated according to the European Society of Hypertension (ESH) International Protocol Revision 2010 in four separate studies. METHODS The four devices measure brachial blood pressure (BP) using the oscillomet...
متن کاملValidation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension
BACKGROUND Four oscillometric devices for self-measurement of blood pressure (SBPM) were evaluated according to the European Society of Hypertension (ESH) international protocol and its 2010 revision in four separate studies. The Omron® M2, Omron M3, and Omron M6 measure blood pressure (BP) at the brachial level, while the Omron R2 measures BP at the wrist level. METHODS The international pro...
متن کاملValidation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population
OBJECTIVE The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010. PARTICIPANTS AND METHODS The evaluation was performed in 39 individuals. The mean age of the participants was 47.9...
متن کاملValidation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
BACKGROUND Allowing patients to measure their blood pressure at home is recognized as being of clinical value. However, it is not known how often these measurements are taken correctly. Blood pressure monitors for home use fall into two types based on the position of the cuff, ie, at the upper arm or the wrist. The latter is particularly convenient, as measurements can be taken fully clothed. T...
متن کاملValidation of the Omron HEM 742 blood pressure monitoring device in adolescents.
BACKGROUND Accurate blood pressure measurement is fundamental for scientific investigation or clinical decision-making. In this sense, it is important to verify the values provided by electronic devices. OBJECTIVE To validate the Omron HEM 742 blood pressure monitoring device in adolescents according to criteria suggested by the British Hypertension Society. METHODS A total of 150 adolescen...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Blood pressure monitoring
دوره 11 3 شماره
صفحات -
تاریخ انتشار 2006