X-ray imaging with amorphous selenium: theoretical feasibility of the liquid crystal light valve for radiography.
نویسندگان
چکیده
A novel radiographic imaging system based on a liquid crystal light valve is described. A liquid crystal light valve is a photon addressed spatial light modulator that consists of a high resolution, solid-state electrostatic detector (photoconductor) and an electro-optic light modulator (liquid crystal cell) physically coupled in a sandwich structure. We propose a light valve with a thick, x-ray sensitive photoconductive layer and call the system under study the x-ray light valve (XLV). The image formation in the XLV is based on x-ray exposure controlled modulation of light from an external source; the XLV is essentially an x-ray image intensifier that allows the image brightness to be adjusted independently from the x-ray exposure. Thus the XLV may be coupled to an optical imager, such as a charge coupled device (CCD), for image digitization without a secondary quantum sink. A model of the XLV operation is developed to investigate its sensitivity, speed, noise, and resolution. The imaging properties of the XLV are found to be time dependent, which leads to an unusual transmission versus exposure characteristic. The feasibility of clinical use of the XLV based on amorphous selenium (a-Se) photoconductor and a twisted nematic liquid crystal cell is analyzed, and the device is shown to be adaptable to a variety of radiographic imaging tasks.
منابع مشابه
The x-ray light valve: a potentially low-cost, digital radiographic imaging system--a liquid crystal cell design for chest radiography.
Digital x-ray radiographic systems are desirable as they offer high quality images which can be processed, transferred, and stored without secondary steps. However, current clinical systems are extraordinarily expensive in comparison to film-based systems. Thus, there is a need for an economical digital imaging system for general radiology. The x-ray light valve (XLV) is a novel digital x-ray d...
متن کاملThe x-ray light valve: a potentially low-cost, digital radiographic imaging system-concept and implementation considerations.
New x-ray radiographic systems based on large-area flat-panel technology have revolutionized our capability to produce digital x-ray images. However, these imagers are extraordinarily expensive compared to the systems they are replacing. Hence, there is a need for a low-cost digital imaging system for general applications in radiology. A novel potentially low-cost radiographic imaging system ba...
متن کاملX-ray detectors for digital radiography.
Digital radiography offers the potential of improved image quality as well as providing opportunities for advances in medical image management, computer-aided diagnosis and teleradiology. Image quality is intimately linked to the precise and accurate acquisition of information from the x-ray beam transmitted by the patient, i.e. to the performance of the x-ray detector. Detectors for digital ra...
متن کاملX-ray imaging using amorphous selenium: determination of Swank factor by pulse height spectroscopy.
The use of photoconductors, especially amorphous selenium (a-Se), in x-ray imaging is currently of interest. A critical performance parameter of an imaging detector is the Swank factor for degradation of the signal to noise ratio, or DQE(0), due to variations in the detector response. The Swank factor is evaluated from measured pulse height spectra generated by the absorption of monoenergetic x...
متن کاملComparison of Pulsed X-ray Source Digital Radiography with Isotopic Radiography on Pipe
Southwest Research Institute (SwRI ® ) conducted a study to determine whether pulsed x-ray sources used in conjunction with amorphous silicon real-time imaging plates could provide equivalent radiographs to those obtained using isotopic sources with conventional radiographic film for inspection of schedule 40 steel pipe. Radiographic procedures for the two techniques were developed; test welds ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 24 8 شماره
صفحات -
تاریخ انتشار 1997