Image-based ghost correction for interleaved EPI.
نویسندگان
چکیده
A new image-based ghost correction technique is described for general interleaved EPI. This technique works reliably with both even- and odd-number interleaved EPI sequences. It estimates the phase distortions causing the ghosts as a general function of (x,y) even in the presence of a significant overlap of parent image and ghosts. If the phase distortion is assumed to be a function of x (frequency-encode direction) only, the new technique is similar to a recently published correction method for general interleaved EPI (Hennel. J Magn Reson 1998;134:206-213). However, that study concluded that the method for general interleaved EPI did not work. It showed that the SNRs of the edge pixels were too low to accurately estimate the phase distortion. The new technique utilizes not one, but many pixels in the image for estimating the phase distortion. The technique requires a user-defined ROI to outline the parent image and identify so-called eligible pixels for estimating the phase distortion. When all eligible pixels are used, the SNR of the phase distortion estimate is comparable to that obtained in single-shot EPI. This study provides the first examples of high-quality, image-based correction of multiple ghosts in general interleaved EPI. Magn Reson Med 45:96-108, 2001.
منابع مشابه
Image-based ghost correction for general interleaved EPI
Introduction A new image-based ghost correction technique is described for general interleaved EPI. It estimates the phase distortions causing the ghosts as a general function of (x,y), even in the presence of significant overlap of parent image and ghosts. If the phase distortion is assumed to be a function of x (frequency encode direction) only, the new technique is similar to a recently publ...
متن کاملDigital Tuning to Suppress Ghost Artifacts in EPI by Minimization of Total Variation
Introduction Echo-planar imaging (EPI) is widely used in functional MRI studies, most often in single-shot form. Interleaved multi-segment EPI becomes desirable when higher spatial resolution and less geometric distortion are crucial. However, interleaved EPI is more susceptible to ghost artifacts. The first major cause of ghost artifacts is even-odd echo data discrepancy, mainly due to gradien...
متن کاملPhased array ghost elimination.
Parallel imaging may be applied to cancel ghosts caused by a variety of distortion mechanisms, including distortions such as off-resonance or local flow, which are space variant. Phased array combining coefficients may be calculated that null ghost artifacts at known locations based on a constrained optimization, which optimizes SNR subject to the nulling constraint. The resultant phased array ...
متن کاملImage-based EPI ghost artifact reduction using iterative phase correction
Abstract. We describe an iterative algorithm for reducing EPI ghost artifacts using image-based phase corrections. Ghosts outside the parent object are set to the background level and a model k-space is obtained from the Fourier transform of the resulting image. The phase offsets in each k-space line are estimated by comparison with the corresponding line in the model k-space. We hypothesise th...
متن کاملHigh spatial resolution EPI using an odd number of interleaves.
Ghost artifacts in echoplanar imaging (EPI) arise from phase errors caused by differences in eddy currents and gradient ramping during left-to-right traversal of kx(forward echo) versus right-to-left traversal of kx (reverse echo). Reference scans do not always reduce the artifact and may make image quality worse. To eliminate the need for reference scans, a ghost artifact reduction technique b...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Magnetic resonance in medicine
دوره 45 1 شماره
صفحات -
تاریخ انتشار 2001