An investigation of a sinogram discontinuity artifact on myocardial perfusion imaging.

نویسندگان

  • Subhash Chand Kheruka
  • Umesh Chand Naithani
  • Lalit Mohan Aggarwal
  • Nirmal Kumar Painuly
  • Anil Kumar Maurya
  • Sanjay Gambhir
چکیده

UNLABELLED Our purpose was to find out the cause of an unusual distinct break seen on a patient's sinogram despite within-tolerance results on all quality assurance tests during myocardial perfusion imaging (MPI) and corrective measures. METHODS SPECT quality control is a prerequisite to obtaining high-quality diagnostic images. Daily uniformity and energy-peaking tests and weekly center-of-rotation (COR) tests are run to check the performance of the SPECT system. A distinct break in the sinogram of an MPI study was noticed for 1 patient, despite routine quality control tests that showed the system to be well within tolerance limits. Critical inspection of the γ-camera revealed that the 2 detector heads did not make complete contact with each other at a 90° position and that a gap of as much as 10.0 mm was left between the 2 edges of the detectors even though COR testing showed the system to be within tolerance limits. After this gap had been minimized (-10-mm position), the MPI study of this patient was repeated. RESULTS Reduction of the gap between detectors corrected the sinogram discontinuity. On the MPI study, the break in the sinogram existed because the 2 detectors were not acquiring the data at the same position in their useful fields of view. When one of the detectors was tilted to exactly 45°, the gap was reduced and the data were acquired at the same useful field of view for both detectors. CONCLUSION The sinogram artifact may arise even after perfect COR calibration, and in the reported case, the discontinuity in the sinogram was rectified by correcting the angle of the detectors. Meticulous investigation for artifacts must be performed to minimize the probability of false results.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The truncation artifact in a skinny patient following myocardial perfusion SPECT

Myocardial perfusion SPECT is one of the most common imaging techniques performed in nuclear medicine departments. To avoid misleading interpretation, it is necessary to address the quality control and technical problems. The truncation artifact occurs when the patient size is large relative to the field of view of the camera, causing false perfusion defects in the LV myocardium, misinterpreted...

متن کامل

Translation of Methdology used in Human Myocardial Imaging to a Sheep Model of Acute Myocardial Infarction

Background: Pre-clinical investigation of stem cells for repairing damaged myocardium predominantly used rodents, however large animals have cardiac circulation closely resembling the human heart. The aim of this study was to evaluate whether SPECT/CT myocardial perfusion imaging (MPI) could be used for assessing sheep myocardium following an acute myocardial infarction (MI) and response to int...

متن کامل

Attenuation correction in myocardial perfusion SPECT using sequential transmission - emission scanning with 99mTc [Persain]

Introduction: Nowadays, Imaging of the myocardial perfusion (MPI) using the single photon emission tomography (SPET) in the diagnosis of coronary artery disease, especially myocardial ischemia, is of great importance. In contrast to the coronary artery angiography, MPI is non-invasive, less expensive and more physiological. Unfortunately, this image is affected by the some artifacts. Thes...

متن کامل

A New Method for Metal Artifact Reduction in CT Scan Images

Introduction In CT imaging, metallic implants inside the tissues cause metal artifact that reduce the quality of image for diagnosis. In order to reduce the effect of this artifact, a new method with more appropriate results has been presented in this research work. Materials and Methods The presented method comprised of following steps: a) image enhancement and metal areas extraction, b) sinog...

متن کامل

Tracing ischemic memory by metabolic pathways: BMIPP and beyond

Myocardial ischemia (MI) resulting in infarction is an important cause of mortality and morbidity worldwide. Acute ischaemia rapidly impairs myocardial contractile function. Myocardial dysfunction persisting for several hours after transient non-lethal ischaemia, eventually resulting in full functional recovery is termed as myocardial stunning. Hibernation is now thought to be...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of nuclear medicine technology

دوره 40 1  شماره 

صفحات  -

تاریخ انتشار 2012