Knotless anchors with sutures external to the anchor body may be at risk for suture cutting through osteopenic bone

نویسندگان

  • Y. Ono
  • J. M. Woodmass
  • A. A. Nelson
  • R. S. Boorman
  • G. M. Thornton
  • I. K. Y. Lo
چکیده

OBJECTIVES This study evaluated the mechanical performance, under low-load cyclic loading, of two different knotless suture anchor designs: sutures completely internal to the anchor body (SpeedScrew) and sutures external to the anchor body and adjacent to bone (MultiFIX P). METHODS Using standard suture loops pulled in-line with the rotator cuff (approximately 60°), anchors were tested in cadaveric bone and foam blocks representing normal to osteopenic bone. Mechanical testing included preloading to 10 N and cyclic loading for 500 cycles from 10 N to 60 N at 60 mm/min. The parameters evaluated were initial displacement, cyclic displacement and number of cycles and load at 3 mm displacement relative to preload. Video recording throughout testing documented the predominant source of suture displacement and the distance of 'suture cutting through bone'. RESULTS In cadaveric bone and foam blocks, MultiFIX P anchors had significantly greater initial displacement, and lower number of cycles and lower load at 3 mm displacement than SpeedScrew anchors. Video analysis revealed 'suture cutting through bone' as the predominant source of suture displacement in cadaveric bone (qualitative) and greater 'suture cutting through bone' comparing MultiFIX P with SpeedScrew anchors in foam blocks (quantitative). The greater suture displacement in MultiFIX P anchors was predominantly from suture cutting through bone, which was enhanced in an osteopenic bone model. CONCLUSIONS Anchors with sutures external to the anchor body are at risk for suture cutting through bone since the suture eyelet is at the distal tip of the implant and the suture directly abrades against the bone edge during cyclic loading. Suture cutting through bone may be a significant source of fixation failure, particularly in osteopenic bone.Cite this article: Y. Ono, J. M. Woodmass, A. A. Nelson, R. S. Boorman, G. M. Thornton, I. K. Y. Lo. Knotless anchors with sutures external to the anchor body may be at risk for suture cutting through osteopenic bone. Bone Joint Res 2016;5:269-275. DOI: 10.1302/2046-3758.56.2000535.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Suture locking of isolated internal locking knotless suture anchors is not affected by bone quality

PURPOSE The purpose of this study was to evaluate the mechanical performance of different suture locking mechanisms including: i) interference fit between the anchor and the bone (eg, 4.5 mm PushLock, 5.5 mm SwiveLock), ii) internal locking mechanism within the anchor itself (eg, 5.5 mm SpeedScrew), or iii) a combination of interference fit and internal locking (eg, 4.5 mm MultiFIX P, 5.5 mm Mu...

متن کامل

Pullout strength of knotless suture anchors.

PURPOSE Suture anchors are used consistently for repairs of soft tissues, especially around the glenohumeral joint. These anchors can be used either arthroscopically or in an open procedure to anatomically restore the labrum and capsular tissues to the glenoid after avulsion injuries (Bankart lesion). The purpose of this study was to analyze the pullout strength of a new knotless suture anchor ...

متن کامل

Quadriceps Tendon Repair Using Knotless Anchors and Suture Tape

Repair of a torn quadriceps tendon is necessary to restore the extensor mechanism of the knee. Traditional repair involves transosseous sutures tied over bone bridges on the inferior pole of the patella. Suture anchor repair has been shown to be stronger than transosseous repair and facilitates a smaller incision. Suture tape can improve the strength of the suture-tendon interface, and when com...

متن کامل

Comparison of the Fixation Strength of PEEK Knotless Suture Anchors for Rotator Cuff Repair

INTRODUCTION Rotator cuff injuries are common, with approximately 400,000 cuff repairs performed each year in the United States alone. Surgical repair of torn cuff tendons involves reattachment of the torn end of the tendon to the original footprint on the proximal humerus, typically using suture anchors. In recent years, suture anchors have evolved into knotless systems, which are designed to ...

متن کامل

Failure mode of suture anchors as a function of insertion depth.

BACKGROUND Surgeons can control not only the angle but also the depth of suture anchor placement during arthroscopic rotator cuff repair, although the tendency may be to place suture anchors on the deep side to avoid damage from prominent anchor eyelets. However, little information is available regarding possible effects of suture anchor depth on construct failure mechanisms. HYPOTHESIS Ancho...

متن کامل