Radiation dose from single-heartbeat coronary CT angiography performed with a 320-detector row volume scanner.
نویسندگان
چکیده
PURPOSE To determine radiation doses from coronary computed tomographic (CT) angiography performed by using a 320-detector row volume scanner and evaluate how the effective dose depends on scan mode and the calculation method used. MATERIALS AND METHODS Radiation doses from coronary CT angiography performed by using a volume scanner were determined by using metal-oxide-semiconductor field-effect transistor detectors positioned in an anthropomorphic phantom physically and radiographically simulating a male or female human. Organ and effective doses were determined for six scan modes, including both 64-row helical and 280-row volume scans. Effective doses were compared with estimates based on the method most commonly used in clinical literature: multiplying dose-length product (DLP) by a general conversion coefficient (0.017 or 0.014 mSv.mGy(-1).cm(-1)), determined from Monte Carlo simulations of chest CT by using single-section scanners and previous tissue-weighting factors. RESULTS Effective dose was reduced by up to 91% with volume scanning relative to helical scanning, with similar image noise. Effective dose, determined by using International Commission on Radiological Protection publication 103 tissue-weighting factors, was 8.2 mSv, using volume scanning with exposure permitting a wide reconstruction window, 5.8 mSv with optimized exposure and 4.4 mSv for optimized 100-kVp scanning. Estimating effective dose with a chest conversion coefficient resulted in a dose as low as 1.8 mSv, substantially underestimating effective dose for both volume and helical coronary CT angiography. CONCLUSION Volume scanning markedly decreases coronary CT angiography radiation doses compared with those at helical scanning. When conversion coefficients are used to estimate effective dose from DLP, they should be appropriate for the scanner and scan mode used and reflect current tissue-weighting factors. (c) RSNA, 2010.
منابع مشابه
Submillisievert median radiation dose for coronary angiography with a second-generation 320-detector row CT scanner in 107 consecutive patients.
PURPOSE To (a) use a new second-generation wide-volume 320-detector row computed tomographic (CT) scanner to explore optimization of radiation exposure in coronary CT angiography in an unselected and consecutive cohort of patients referred for clinical purposes and (b) compare estimated radiation exposure and image quality with that from a cohort of similar patients who underwent imaging with a...
متن کاملRecent Update on Radiation Dose Assessment for the State-of-the-Art Coronary Computed Tomography Angiography Protocols
OBJECTIVES This study aimed to measure the absorbed doses in selected organs for prospectively ECG-triggered coronary computed tomography angiography (CCTA) using five different generations CT scanners in a female adult anthropomorphic phantom and to estimate the effective dose (HE). MATERIALS AND METHODS Prospectively ECG-triggered CCTA was performed using five commercially available CT scan...
متن کاملCancerogenesis Risks between 64 and 320 Row Detector CT for Coronary CTA Screening
OBJECTIVES This study compares cancerogenesis risks posed by the 64 row detector and the 320 row detector computed tomography scanners used during coronary computed tomography angiography (CCTA) following decennial screening guidelines. MATERIAL AND METHODS Data of the radiation absorbed after CCTA by lung, thyroid, and female breast in patients between 50 and 70 years of age obtained from pr...
متن کاملThe effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography.
OBJECTIVE To evaluate the effect of adaptive iterative dose reduction (AIDR) on image noise and image quality as compared with standard filtered back projection (FBP) in 320-detector row CT coronary angiography (CTCA). METHODS 50 patients (14 females, mean age 68 ± 9 years) who underwent CTCA (100 kV or 120 kV, 400-580 mA) within a single heartbeat were enrolled. Studies were reconstructed wi...
متن کاملCoronary computed tomography angiography with 320-row detector and using the AIDR-3D: initial experience
Coronary computed tomography angiography (coronary CTA) is a powerful non-invasive imaging method to evaluate coronary artery disease. Nowadays, coronary CTA estimated effective radiation dose can be dramatically reduced using state-of-the-art scanners, such as 320-row detector CT (320-CT), without changing coronary CTA diagnostic accuracy. To optimize and further reduce the radiation dose, new...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Radiology
دوره 254 3 شماره
صفحات -
تاریخ انتشار 2010