Direct Measurement of Mammographic X-Ray Spectra with a Digital CdTe Detection System

نویسندگان

  • Leonardo Abbene
  • Gaetano Gerardi
  • Fabio Principato
  • Stefano Del Sordo
  • Giuseppe Raso
چکیده

In this work we present a detection system, based on a CdTe detector and an innovative digital pulse processing (DPP) system, for high-rate X-ray spectroscopy in mammography (1-30 keV). The DPP system performs a height and shape analysis of the detector pulses, sampled and digitized by a 14-bit, 100 MHz ADC. We show the results of the characterization of the detection system both at low and high photon counting rates by using monoenergetic X-ray sources and a nonclinical X-ray tube. The detection system exhibits excellent performance up to 830 kcps with an energy resolution of 4.5% FWHM at 22.1 keV. Direct measurements of clinical molybdenum X-ray spectra were carried out by using a pinhole collimator and a custom alignment device. A comparison with the attenuation curves and the half value layer values, obtained from the measured and simulated spectra, from an ionization chamber and from a solid state dosimeter, also shows the accuracy of the measurements. These results make the proposed detection system a very attractive tool for both laboratory research, calibration of dosimeters and advanced quality controls in mammography.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Energy, angular and spatial distributions of primary electrons inside photoconducting materials for digital mammography: Monte Carlo simulation studies.

Materials such as a-Se, a-As(2)Se(3), GaSe, GaAs, Ge, CdTe, CdZnTe, Cd(0.8)Zn(0.2)Te, ZnTe, PbO, TlBr, PbI(2) and HgI(2) are potential candidates as photoconductors in direct detectors for digital mammography. The x-ray induced primary electrons inside a photoconductor's bulk comprise the initial signal that propagates and forms the final signal (image) on the detector's electrodes. An already ...

متن کامل

X-ray spectral measurements for tungsten-anode from 20 to 49 kVp on a digital breast tomosynthesis system.

PURPOSE This paper presents new spectral measurements of a tungsten-target digital breast tomosynthesis (DBT) system, including spectra of 43-49 kVp. METHODS Raw x-ray spectra of 20-49 kVp were directly measured from the tube port of a Selenia Dimensions DBT system using a CdTe based spectrometer. Two configurations of collimation were employed: one with two tungsten pinholes of 25 μm and 200...

متن کامل

Calculation of the X-Ray Spectrum of a Mammography System with Various Voltages and Different Anode-Filter Combinations Using MCNP Code

Introduction One of the best methods in the diagnosis and control of breast cancer is mammography. The importance of mammography is directly related to its value in the detection of breast cancer in the early stages, which leads to a more effective treatment. The purpose of this article was to calculate the X-ray spectrum in a mammography system with Monte Carlo codes, including MCNPX and MCNP5...

متن کامل

Calculation of Absorbed Glandular Dose using a FORTRAN Program Based on Monte Carlo X-ray Spectra in Mammography

Introduction: Average glandular dose calculation in mammography with Mo-Rh target-filter and dose calculation for different situations is accurate and fast. Material and Methods: In this research, first of all, x-ray spectra of a Mo target bombarded by a 28 keV electron beam with and without a Rh filter were calculated using the MCNP code. Then, we used the Sobol-Wu parameters to write a FORTRA...

متن کامل

MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA USING CdZnTe DETECTOR

INTRODUCTION We directly measured diagnostic X-ray photon spectra produced by a single-phase 2-pulse diagnostic X-ray unit driven at 50 to 100 kV and a few mA(1), a mammographic X-ray unit at 25 to 32 kV and 3 times of 80 mAs(2) and an X-ray computed tomography (CT) scanner at 120 kV and 3 mA(3) have been analyzed. Since detected spectra were distorted by the response of CdZnTe(CZT) detector an...

متن کامل