SU-E-T-453: Optimization of Dose Gradient for Gamma Knife Radiosurgery.

نویسندگان

  • N Sheth
  • Y Chen
  • J Yang
چکیده

PURPOSE The goals of stereotactic radiosurgery (SRS) are the ablation of target tissue and sparing of critical normal tissue. We develop tools to aid in the selection of collimation and prescription (Rx) isodose line to optimize the dose gradient for single isocenter intracranial stereotactic radiosurgery (SRS) with GammaKnife 4C utilizing the updated physics data in GammaPlan v10.1. METHODS Single isocenter intracranial SRS plans were created to treat the center of a solid water anthropomorphism head phantom for each GammaKnife collimator (4 mm, 8 mm, 14 mm, and 18 mm). The dose gradient, defined as the difference of effective radii of spheres equal to half and full Rx volumes, and Rx treatment volume was analyzed for isodoses from 99% to 20% of Rx. RESULTS The dosimetric data on Rx volume and dose gradient vs. Rx isodose for each collimator was compiled into an easy to read nomogram as well as plotted graphically. The 4, 8, 14, and 18 mm collimators have the sharpest dose gradient at the 64%, 70%, 76%, and 77% Rx isodose lines, respectively. This corresponds to treating 4.77 mm, 8.86 mm, 14.78 mm, and 18.77 mm diameter targets with dose gradients radii of 1.06 mm, 1.63 mm, 2.54 mm, and 3.17 mm, respectively. CONCLUSION We analyzed the dosimetric data for the most recent version of GammaPlan treatment planning software to develop tools that when applied clinically will aid in the selection of a collimator and Rx isodose line for optimal dose gradient and target coverage for single isocenter intracranial SRS with GammaKnife 4C.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Long term effects of Gamma knife Radiosurgery for treatment of cerebral arteriovenous malformations

 Abstract Background: The Gamma Knife Radiosurgery (GKR) is an established management option for Cerebral Ar-teriovenous Malformations (AVMS). Therapeutic benefits of radiosurgery for arteriovenous malformations are complete obliteration of nidus with minimal neurological deficit. Methods: Radiosurgery was performed between February 2003 and April 2010 at Kamraniye day clinic, Teh-ran, Iran, us...

متن کامل

SU-E-T-445: A Dosimetric Comparison of Gamma Knife Plans with Measurement Based and Image Based Skull Definitions.

PURPOSE To compare gamma knife treatment plans that use skull scaling instrument measurements for skull definition vs. CT based skull definition. METHODS Eight previously treated patients who had head CT scans were randomly selected. Their CT images were imported into Leksell GammaPlan. For each patient, three dose plans were developed: 1. the reference plan. It was planned based on the stand...

متن کامل

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...

متن کامل

Evaluation of Organ Specific peripheral dose for Gamma knife 4C based on Monte Carlo

Introduction: Stereotactic Gamma Knife radiosurgery has been widely used for treating brain tumors. The scattered radiation outside of treatment field (peripheral dose) can induce the secondary cancer to specific organ. This paper investigated the absorbed dose to eyes, thyroid, heart, lung, breast and colon using a Monte Carlo technique for Mird phantom. We also study the ef...

متن کامل

Comparison dose distributions from gamma knife unit 4C with CT data and non-CT data options of beamnrc code

Todays gamma knife radiosurgery is used widely for treatment of very small brain tumors. In order to investigate accuracy of dosimetry and treatment planning calculations, using Monte Carlo simulation with dedicated code named as beamnrc including non-CT data and CT data options is necessary. The aim of this study is choosing the best options in order to have an accurate tools based on their ad...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 39 6Part17  شماره 

صفحات  -

تاریخ انتشار 2012