Characterization of a ZnSe scintillating bolometer prototype for neutrinoless double beta decay search
نویسندگان
چکیده
As proposed in the LUCIFER project, ZnSe crystals are attractive materials to realize scintillating bolometers aiming at the search for neutrinoless double beta decay of the promising isotope 82Se. However, the optimization of the ZnSe-based detectors is rather complex and requires a wide-range investigation of the crystal features: optical properties, crystalline quality, scintillation yields and bolometric behaviour. Samples tested up to now show problems in the reproducibility of crucial aspects of the detector performance. In this work, we present the results obtained with a scintillating bolometer operated aboveground at about 25 mK. The detector energy absorber was a single 1 cm3 ZnSe crystal. The good energy resolution of the heat channel (about 14 keV at 1460 keV) and the excellent alpha/beta discrimination capability are very encouraging for a successful realization of the LUCIFER program. The bolometric measurements were completed by optical tests on the crystal (optical transmission and luminescence measurements down to 10 K) and investigation of the crystalline structure. The work here described provides a set of parameters and procedures useful for a complete pre-characterization of ZnSe crystals in view of the realization of highly performing scintillating bolometers.
منابع مشابه
An experimental study of antireflective coatings in Ge light detectors for scintillating bolometers
Luminescent bolometers are double-readout devices able to measure simultaneously the phonon and the light yields after a particle interaction in the detector. This operation allows in some cases to tag the type of the interacting quantum, crucial issue for background control in rare event experiments such as the search for neutrinoless double beta decay and for interactions of particle dark mat...
متن کاملLi-containing scintillating bolometers for low background physics
We present the performances of Li-based compounds used as scintillating bolometer for rare decay studies such as double-beta decay and direct dark matter investigations. The compounds are tested in a dilution refrigerator installed in the underground laboratory of Laboratori Nazionali del Gran Sasso (Italy). Low temperature scintillating properties are investigated by means of different radioac...
متن کاملStatus of Cuore experiment and last results from Cuoricino
CUORE is a cryogenic-bolometer detector consisting of 988 TeO 2 crystals, 750 g each, operated at a temperature of ∼8 mK, currently under construction in the underground Laboratori Nazionali del Gran Sasso. Its goal is to search for neutrinoless double beta decays with a sensitivity to the effective neutrino mass as low as a few tens of meV. CUORICINO, its pilot experiment, has proven the feasi...
متن کاملLUMINEU: a search for neutrinoless double beta decay based on ZnMoO4 scintillating bolometers
The LUMINEU is designed to investigate the possibility to search for neutrinoless double beta decay in Mo by means of a large array of scintillating bolometers based on ZnMoO4 crystals enriched in Mo. High energy resolution and relatively fast detectors, which are able to measure both the light and the heat generated upon the interaction of a particle in a crystal, are very promising for the re...
متن کاملScintillating double beta decay bolometers
P. Gorla Laboratori Nazionali del Gran Sasso, I-67010, Assergi (L’Aquila), Italy Abstract We present the results obtained in the development of scintillating Double Beta Decay bolometers. Several Mo and Cd based crystals were tested with the bolometric technique. The scintillation light was measured through a second independent bolometer. A 140 g CdWO4 crystal was run in a 417 h live time measu...