Performance of a radiometer transcutaneous oxygen monitor in a neonatal-intensive-care unit.
نویسندگان
چکیده
We evaluated the performance of the Radiometer TCM-1 Transcutaneous Oxygen Monitor in a Class III neonatal-intensive-care unit during an 11-month clinical trial (in excess of 4000 h) in which we monitored 115 infants ranging in gestational age from 25 to 43 weeks and in birthweight from 595 to 4220 g. In vitro transcutaneous partial pressure of oxygen (ptcO2) measurements were accurate within 1% of the theoretical value and precise (1.4% CV at 156 mmHg) at 44 degrees C. Determinations at other temperature were less accurate except in rare instances when specific patient-associated ptcO2/paO2 bias was identified. In vitro comparison of ptcO2 (y) and arterial partial pressure of oxygen (paO2) (x) measurements agreed well (y = 0.90x + 6.9, r = 0.94) except when infants were less than 24 h old. In this latter group the ptcO2/paO2 ratio usually exceeded 1.0 and fluctuated unpredictably. ptcO2 measurements on infants less than 24 h old had to be interpreted cautiously. In general, the ptcO2 and paO2 measurements agreed sufficiently well for the physicians to use them as the basis for directing patient care.
منابع مشابه
The Effect of Non-nutritive Sucking on Transcutaneous Oxygen Saturation in Neonates under the Nasal Continuous Positive Airway Pressure (CPAP)
BackgroundSeveral beneficial effects of non-nutritive sucking in infants, including the physiological stability, relaxation, better transition from tube feeding to oral feeding have been reported. But its effect on oxygen saturation in neonates under the Nasal Continuous Positive Airway Pressure (NCPAPو (is not so clear. This study aimed to investigate the effects of non-nutritive sucking on tr...
متن کاملUse of a Combined SpO2/PtcCO2 Sensor in the Delivery Room
Arterial oxygen saturation (SaO(2)) and partial arterial pressure of carbon dioxide (PaCO(2)) are important respiratory parameters in critically ill neonates. A sensor combining a pulse oximeter with the Stow-Severinghaus electrode, required for the measurement of peripheral oxygen saturation (SpO(2)) and transcutaneous partial pressure of carbon dioxide (PtcCO(2)), respectively, has been recen...
متن کاملAssessing the Performance of Nurses in the Proper Adjustment of Monitoring Instruments in the Neonatal Intensive Care Unit
Background: Monitoring serves to maintain physiologic variables within normal limits and when a parameter crosses a set threshold, an alarm is triggered. Therefore, this study aimed to determine whether alarm limits are properly adjusted in the neonatal intensive care unit (NICU) by nursing staff or not.Methods: The data concerning alarm limits corresponding to the three shifts of nursing...
متن کاملPrevalence of Hearing Loss in Newborns Admitted to Neonatal Intensive Care Unit
Introduction: Hearing is essential for humans to communicate with one another. Early diagnosis of hearing loss and intervention in neonates and infants can reduce developmental problems. The aim of the present study was to assess the prevalence of hearing impairment in newborns admitted to a neonatal intensive care unit (NICU) and analyze the associated risk factors. Materials and Methods:...
متن کاملPerformance of Neonatal Intensive Care Unit Nurses in Blood Culture Procedure in Tabriz Hospitals in 2016
Background: Blood culture is a method in neonatal intensive care unit (NICU) and a gold standard for the diagnosis of infection that is essential for proper newborn management. The aim of this study was to compare the performance of nurses in the NICUs of Tabriz University of Medical Sciences using a standard sampling method. Methods: This descriptive cross-sectional study was performed on 90 ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Clinical chemistry
دوره 26 5 شماره
صفحات -
تاریخ انتشار 1980