Effect of solvent flow rate in mini-column testing of 99mTc-mertiatide.

نویسنده

  • James A Ponto
چکیده

OBJECTIVE The recommended method for radiochemical purity testing of 99mTc-mertiatide involves the use of a C-18 solid-phase mini-column cartridge. The mertiatide package insert states that the solvents should be "pushed through the cartridge slowly," but a flow rate is not specified. The mini-column cartridge instruction sheet recommends flow rates of 5-10 and 2-10 mL/min for conditioning and for elution, respectively, of the cartridge. The purpose of this study was to evaluate the effect of different flow rates on determining the radiochemical purity of 99mTc-mertiatide. METHODS Radiochemical purity was tested on 10 consecutive vials of 99mTc-mertiatide prepared for routine clinical use and on 4 vials of 99mTc-mertiatide spiked with 6%-15% free pertechnetate using 3 different flow rates: slow drip (5 mL/min for conditioning and 2 mL/min for elution), fast drip (10 mL/min for conditioning and 10 mL/min for elution), and very fast drip (about 15-20 mL/min for conditioning and about 15-20 mL/min for elution). An infusion pump was used to provide constant flow rates for the first 2 conditions, whereas manual handling, reflecting real-life practice, was used for the third condition. RESULTS All 3 flow rates yielded essentially identical radiochemical purities for each vial tested (agreement was always within 0.3% for a given vial). The elapsed times for mini-column conditioning, loading, and elution were approximately 15, 5, and 3 min for the slow drip, fast drip, and very fast drip, respectively. CONCLUSION Faster flow rates for mini-column testing of 99mTc-mertiatide save time (and correspondingly reduce radiation exposure to the worker) without adversely affecting the results of radiochemical purity determinations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Optimization of Atropine Extraction Process from Atropa Belladonna by Modified Bubble Column Extractor with Ultrasonic Bath

Modified Bubble Column Extraction with Ultrasonic Bath (BCE-UB) method was used to extract atropine from the stem and leaves of Atropa belladonna. Optimum condition were obtained with Kamada solvent which was chloroform-methanol-ammonia 15:15:1(v/v/v) as extraction solvent, the particle size of  less than  350 µm, an extraction time of 23.95 min, a liquor to material ratio of 15.08 mL/g and...

متن کامل

Preparation, radiochemical purity control and stability of 99mTc-mertiatide (Mag-3).

BACKGROUND Scintigraphic image analysis of 99mTc-mertiatide (Mag-3, mercaptoacetyltriglycine) clearance provides the determination of the blood flow, the tubular transit time and the excretion as well from both kidneys. Radiopharmaceutical routine recommends a radiochemical purity control before administration of the product to a patient. The main objective of this study is to develop a Mag-3 l...

متن کامل

Renal artery stenosis and ischemia. Effect on renal blood flow and extraction fraction.

The clearance of [131I]orthoiodohippurate and 99mTc-mercaptoacetyltriglycine provide a measure of effective renal plasma flow, yet these clearances are proportional to renal plasma flow only if the extraction fraction remains constant. To determine the effect of unilateral renal artery stenosis, captopril, renal ischemia, and partial renal vein occlusion on renal blood flow and the extraction f...

متن کامل

بهینه سازی روش تغلیظ و تخلیص ترانس، ترانس - موکونیک اسید به عنوان شاخص بیولوژیکی بنزن در مواجهه های شغلی

In this study, parameters affecting on SPE of trance trance muconic acid (t,t-MA), including sample pH, sample concentration, sample volume, sample flow rate, washing solvent, elution solvent, and type of sorbent were evaluated. After sample preparation of muconic acid as urinary metabolite of benzene, it was determined by HPLC-UV. In chromatographic analysis, column was C18 (250 cm × 4.6 mm id...

متن کامل

Experimental and Computational Study on Hydrodynamic of a Downscaled Mini Vessel USP Dissolution Test Apparatus II

Although not listed on the United States Pharmacopeia (USP), like standard USP 2, small volume USP 2 dissolution apparatus has gained a great deal of attention, especially for cases where small amount of drug product is available for testing in research and design step or evaluations are to be made on a tablet containing trace amounts of the active pharmaceutical ingredient. In this work, first...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of nuclear medicine technology

دوره 33 4  شماره 

صفحات  -

تاریخ انتشار 2005