Wire Scanners for Small Emittance Beam Measurement in Atf
نویسنده
چکیده
The wire scanners are used for a measurement of the very small beam size and the emittance in Accelerator Test Facility (ATF). They are installed in the extraction beam line of ATF damping ring. The extracted beam emittance are εx=1.3x10 m.rad, εy=1.7x10 m.rad with 2x10 electrons/bunch intensity and 1.3GeV energy[1]. The wire scanners scan the beam by a tungsten wire with beam repetition 0.78Hz. The scanning speed is, however, very slow( ~500μm/sec). Since the extracted beam is quite stable by using the double kicker system[2], precision of the size measurement is less than 2μm for 50 150μm horizontal beam size and 0.3μm for 8 16μm vertical beam size. The detail of the system and the performance are described.
منابع مشابه
4d Emittance Measurements Using Multiple Wire and Waist Scan Methods in the Atf Extraction Line
Emittance measurements performed in the diagnostic section of the ATF extraction line since 1998 lead to vertical emittances three times larger than the expected ones, with a strong dependence on intensity. An experimental program is pursued to investigate potential sources of emittance growth and find possible remedies. This requires efficient and reliable emittance measurement techniques. In ...
متن کاملStatus of the ATF extraction line laser-wire
The Accelerator Test Facility (ATF) at KEK offers a very low emittance micrometre sized beam of 1.28 GeV electrons close from the conditions expected at the International Linear Collider (ILC). The ATF should soon be able to provide a bunch spacing similar to that of the ILC. This accelerator being a test facility it is possible to have dedicated machine time with custom beam optics for our tes...
متن کاملSimulation Studies on the Vertical Emittance Growth at the Existing Atf Extraction Beamline*
Significant beam intensity-dependence of the vertical emittance growth was experimentally observed at the Accelerator Test Facility (ATF) at KEK extraction beamline. This paper presents the simulations of possible vertical emittance growth sources, particularly in the extraction channel, where the magnets are shared by both the ATF extraction beamline and its damping ring. The vertical emittanc...
متن کاملImprovements in Emittance Wakefield Optimization for the Slac Linear Collider
The transverse emittances in the SLAC Linear Collider can be severely diluted by collective wakefield effects and dispersion. For the 1997/98 SLC/SLD run important changes were implemented in the way the emittance is optimized. Early in the linac, where the energy spread is large due to BNS damping, the emittance growth is dominated by dispersion. In this regime emittance tuning bumps may intro...
متن کاملIntrabeam Scattering Analysis of Atf Beam Data Taken in April 2000 * Intrabeam Scattering Analysis of Atf Beam Data Taken in April 2000
Our theoretical comparisons suggest that the ATF measurement results of April 2000 for energy spread, bunch length, and horizontal emittance vs. current, and a low current emittance ratio of about 1% are generally consistent with intrabeam scattering (IBS) theory, though the measured effects appear to be stronger than theory. In particular, the measured factor of 3 growth in vertical emittance ...
متن کامل