Effect of epicardial patch electrodes on transthoracic defibrillation.
نویسندگان
چکیده
To improve survival rates in patients undergoing surgical ablative procedures for malignant ventricular tachycardia (VT), a frequent practice is to implant epicardial patch electrodes at the time of map-guided surgery. After operation, patients with inducible VT often receive an automatic internal cardioverter/defibrillator (AICD) implant, whereas patients with noninducible VT usually do not. In the event that spontaneous, hypotensive VT or ventricular fibrillation should subsequently occur in the patient with noninducible VT, however, transthoracic defibrillation and resuscitation may prove difficult, because the patch electrodes are insulated with silicone rubber that can reduce the amount of current traversing the myocardium and thus can increase transthoracic defibrillation threshold (DFT). In this study, DFT was determined in mongrel dogs to test the hypothesis that epicardial patch electrodes elevate threshold. This study was also designed to assess the effect of patch electrode orientation and size on DFT. In the first protocol (perpendicular orientation), small epicardial patch electrodes (surface area, 30 cm2) were sutured to the epicardial surfaces of the anterior right and posterior left ventricles in 15 dogs so that the center axes of the patch electrodes were coincident and perpendicular to the coincident center axes of the transthoracic electrodes. The effect of two large epicardial patch electrodes (surface area, 53 cm2) on transthoracic DFT was also examined in eight of these dogs. In the second protocol (parallel orientation), small patch electrodes were sutured to the lateral surfaces of the right and left ventricles in seven dogs.(ABSTRACT TRUNCATED AT 250 WORDS)
منابع مشابه
Transthoracic Defibrillation
To improve survival rates in patients undergoing surgical ablative procedures for malignant ventricular tachycardia (VT), a frequent practice is to implant epicardial patch electrodes at the time of map-guided surgery. After operation, patients with inducible VT often receive an automatic internal cardioverter/defibrillator (AICD) implant, whereas patients with noninducible VT usually do not. I...
متن کاملComparison of defibrillation probability of success curves for an endocardial lead configuration with and without an inactive epicardial patch.
OBJECTIVES This study sought to assess the effect of passive "bystander" epicardial electrodes on defibrillation efficacy. BACKGROUND We hypothesized that an inactive epicardial patch placed in an area of low potential gradient from an endocardial electrode shock might affect defibrillation efficacy through its effects on the shock field and the underlying potential gradient. METHODS We stu...
متن کاملRelation between transcardiac and transthoracic current during defibrillation in humans.
Conceptually, transthoracic defibrillation threshold current can be considered a function of at least two quantities. It is directly proportional to the transcardiac threshold current and inversely proportional to the transcardiac current fraction (FC) or the ratio of transcardiac and transthoracic current. Although experimental and theoretical estimates of FC have been as high as 45%, previous...
متن کاملIntrathoracic current flow during transthoracic defibrillation in dogs. Transcardiac current fraction.
To achieve transcardiac threshold current during transthoracic defibrillation, a considerably larger current must be delivered to the thorax to compensate for the shunting effect of the lungs, the thoracic cage, and other elements of the torso. This shunting effect is thus an important determinant of transthoracic defibrillation threshold and can be quantified by the transcardiac current fracti...
متن کاملSodium Channel Block and Action Potential Prolongation
Antiarrhythmic drugs have been reported to produce variable effects on defibrillation energy requirements. However, the relation between the in vitro electrophysiologic effects of these agents and the changes in defibrillation energy requirements have not been systematically examined. Therefore, we evaluated the effects of the sodium channel blocking drugs lidocaine and procainamide, the action...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Circulation
دوره 81 4 شماره
صفحات -
تاریخ انتشار 1990