Use of BACTEC MGIT 960 for recovery of mycobacteria from clinical specimens: multicenter study.

نویسندگان

  • E Tortoli
  • P Cichero
  • C Piersimoni
  • M T Simonetti
  • G Gesu
  • D Nista
چکیده

The BACTEC MGIT 960 instrument is a fully automated system that exploits the fluorescence of an oxygen sensor to detect growth of mycobacteria in culture. Its performance was compared to those of the radiometric BACTEC 460 instrument and egg-based Lowenstein-Jensen medium. An identical volume of sample was inoculated in different media, and incubation was carried out for 6 weeks with the automatic systems and for 8 weeks on solid media. A total of 2,567 specimens obtained from 1,631 patients were cultured in parallel. Mycobacteria belonging to nine different taxa were isolated by at least one of the culture systems, with 75% of them being represented by Mycobacterium tuberculosis complex. The best yield was obtained with the BACTEC 460 system, with 201 isolates, in comparison with 190 isolates with the BACTEC MGIT 960 system and 168 isolates with Lowenstein-Jensen medium. A similar but not significant difference was obtained when the most-represented organisms, the M. tuberculosis complex, Mycobacterium xenopi, and the Mycobacterium avium complex, were analyzed separately and when combinations of a solid medium with the BACTEC MGIT 960 system and with the BACTEC 460 system were considered. The shortest times to detection were obtained with the BACTEC MGIT 960 system (13.3 days); 1.5 days earlier than that with the BACTEC 460 system (14.8 days) and 12 days earlier than that with Lowenstein-Jensen medium (25.6 days). The BACTEC MGIT 960 system had a contamination rate of 10.0%, intermediate between those of the radiometric system (3.7%) and the egg-based medium (17.0%). We conclude, therefore, that the BACTEC MGIT 960 system is a fully automated, nonradiometric instrument that is suitable for the detection of growth of tuberculous and other mycobacterial species and that is characterized by detection times that are even shorter than that of the "gold standard," the BACTEC 460 system. The contamination rate was higher than that for the radiometric BACTEC 460 system and needs to be improved.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Multicenter evaluation of the BACTEC MGIT 960 system for recovery of mycobacteria.

We evaluated the BACTEC MGIT 960 system, which is a fully automated, noninvasive system for the growth and detection of mycobacteria with a capacity to incubate and continuously monitor 960 7-ml culture tubes. We studied 3,330 specimens, 2,210 respiratory and 1,120 nonrespiratory specimens, collected from 2,346 patients treated at six sites. Processed specimens were inoculated into the BACTEC M...

متن کامل

Use of the BACTEC Mycobacteria Growth Indicator Tube 960 automated system for recovery of Mycobacteria from 9,558 extrapulmonary specimens, including urine samples.

The BACTEC Mycobacteria Growth Indicator Tube 960 (MGIT 960) system was applied for recovery of mycobacteria from extrapulmonary specimens and compared with solid media (Löwenstein-Jensen and Stonebrink). A total of 9,558 specimens were investigated, comprising 3,074 body fluids, 1,878 tissues, and 2,069 urine samples, from which the recovery of mycobacteria was not yet established for MGIT 960...

متن کامل

Comparison of the BACTEC MGIT 960 and ESP culture system II for growth and detection of mycobacteria.

The performances of two continuously monitoring mycobacterial culture systems-ESP Culture System II (ESP II; Trek Diagnostics, Inc. , Westlake, Ohio) and BACTEC MGIT 960 (BD Biosciences, Sparks, Md. )-were compared. In addition to both liquid media, all specimens were plated onto Middlebrook 7H11/7H11 selective agar. A total of 3, 151 specimens of all types (56.3% were respiratory specimens) we...

متن کامل

Comparison of recoveries of mycobacterium tuberculosis using the automated BACTEC MGIT 960 system, the BACTEC 460 TB system, and Löwenstein-Jensen medium.

Using two different liquid media and one conventional solid medium, a total of 57 mycobacterial isolates (Mycobacterium tuberculosis, n = 55; nontuberculous mycobacteria, n = 2) were recovered from 377 clinical specimens. The rates of recovery of M. tuberculosis were 96. 4% with the BACTEC MGIT 960 liquid medium, 92.7% with BACTEC 12B liquid medium, and 81.8% with the Löwenstein-Jensen (LJ) med...

متن کامل

Evaluation of BACTEC MGIT 960 and BACTEC 460TB systems for recovery of mycobacteria from clinical specimens of a university hospital with low incidence of tuberculosis.

Clinical samples obtained over a period of 8 months (n = 2,624) were processed in parallel with the BACTEC 460TB system, with the MGIT 960 system, and in Löwenstein-Jensen (LJ) medium, resulting in the recovery of 127 mycobacteria. Recovery rates in combinations of the BACTEC 460TB or MGIT 960 system with LJ were, respectively, 94.7 and 94.7% for Mycobacterium tuberculosis complex (n = 57) and ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 37 11  شماره 

صفحات  -

تاریخ انتشار 1999