pTX Spoke RF Pulses for Cardiac MRI at 7T: a New Design Robust against Respiration Induced Errors, based on a Virtual Simultaneous Exhale-and-Inhale Calilbration Scan
نویسندگان
چکیده
INTRODUCTION. Cardiac MRI may strongly benefit from high SNR at ultra-high fields (UHF) but is challenged by the short RF wavelength that induces large flip angle variations over the heart. This problem has been successfully addressed in 7T cardiac CINE imaging using parallel transmission (pTX) with 2-spoke pTX RF pulses [1]. Although 2 spokes can provide high contrast uniformity, we sometimes observed significant deviations between measurement and prediction by Bloch simulation as discussed in [1]. Also, at times, the performance changed when using two different 2-spoke trajectories in k-space with similar prediction (Fig1). We suspected motion to be involved (some trajectories may be more prone than others) but subject’s bulk motion in the scanner was excluded. All calibrations and acquisitions are performed under breath-hold, assuming that subjects always return to the same exhale position. If, however, the subject’s respiratory state (ideally exhaled) during cardiac CINE scans is not the same as during transmit B1 (B1+) and ∆B0 calibration scans, degradation in the performance of the 2-spoke RF pulse uniformity can be expected. Here we investigate the impact of breath-hold position on cardiac CINE imaging using 2-spoke pTX pulses and we demonstrate an RF pulse design strategy to increase the robustness of 2-spoke pTX RF pulse to variations in breath-hold position. Our findings may also impact free-breathing pTX acquisitions, including at 3T. METHODS. After obtaining consent, volunteers were imaged at 7T with a 16 channel pTX prototype console (Siemens, Erlangen) using a 16 channel transceiver body coil [2]. The 16 B1+ sensitivity maps were obtained in an axial plane from ECG triggered GRE scans (1 B1+ map per heartbeat) acquired during a single breath-hold using a fast, small flip angle (FA) calibration [1,3]. For each TX map we also acquired within the same heartbeat a sagittal map (16 in total) crossing the diaphragm to monitor breath-hold position during B1+ calibration. A ∆B0 map, required for 2-spoke RF pulse design, was derived from a dual TE GRE acquisition that also included a diaphragm scan. The B1 + and ∆B0 mapping procedure was repeated 3 times while instructing the subject to hold their breath on 3 different breathhold positions (POS): I) full exhale, II) half inhale, III) full inhale (POS verified with diaphragm data). With slice selection along the z-axis, gradient blips applied along the x and y axes were used to symmetrically position the 2 spokes about kx-ky=0 (see Fig. 2a). In polar coordinates, the radius kr varied from 0 to 10m -1 in 1m steps while the angle φ of the ‘spokes axis’ (dashed line in Fig2a) to the kx axis was varied from 0° to 360° (note: not 180°) in 10° steps. RF pulses were designed as described in [1] based on B1 + and ∆B0 maps acquired in POS I, using a magnitude least squares solution [4] for each of the 361 different trajectories. We performed Bloch simulations for each solution, first with the original B1 + and ∆B0 maps (obtained at POS I), then after substituting the latter with those obtained at POS II and then at POS III. For each spoke trajectory characterized by (kr,φ) and each breath-hold position, we calculated the coefficient of variation (CV= std(FA)/mean(FA)) of the simulated FA profile, as well as the total RF energy (En) (sum of energy through all channels and spokes) and the maximum energy per channel (Emax). This procedure was repeated, but with a modified pulse design based on a virtual 2-slice calibration, including in one slice B1+ and B0 maps from POS-I, and in a second slice B1 +
منابع مشابه
Improved excitation fidelity in cardiac imaging with 2-spoke parallel excitation at 7 Tesla
Background Cardiac MRI may greatly benefit from ultra high field (UHF) providing higher SNR and intrinsic contrast. But shorter RF wavelength yields transmit B1 (B1+) heterogeneity and contrast variations through the heart. These can be addressed by parallel transmission (pTX) using multispoke RF pulses as previously shown in other organs. However, applying pTX in cardiac MRI at 7T is challengi...
متن کاملZ-stacked RF array design enhances parallel transmit multiband RF performance in whole brain simultaneous multislice imaging at 7T
Introduction: Simultaneous MultiSlice (SMS) MR imaging (1) using MultiBand (MB) RF pulses is becoming increasingly popular in the neuroimaging community (2-4). Recently, there has been an interest in utilizing multielement RF arrays in combination with multichannel (pTx) MB pulse design to reduce transmit B1 (B1+) inhomogeneity and SAR (5,6) for SMS/MB imaging. Meanwhile, it has been shown that...
متن کاملWhole brain single shot diffusion weighted EPI at 7 Tesla using parallel transmit multislice multiband RF pulses
Introduction: Simultaneous Multi-Slice (SMS) imaging [1] using MultiBand (MB) RF pulses has shown great success in functional and diffusion weighted (DW) MRI studies of the brain [2-5]. Recently, this technique has been combined with parallel transmission (pTx) and it has been demonstrated that pTx MB pulse design [6] can significantly improve transmit B1 (B1+) homogeneity at 7T and/or reduce R...
متن کاملBreath-holding during the Calibration Scan Improves the Reproducibility of Parallel Transmission at 7T for Human Brain
PURPOSE The B0 and B1+ maps required for calculation of the radiofrequency (RF) pulse of parallel transmission (pTx) are obtained in calibration scans; however, they may be affected by respiratory motion. We aimed to compare the reproducibility of B0 and B1+ maps and gradient echo (GRE) images of the brain scanned with pTx at 7T between free-breathing (FB) and breath-holding (BH) conditions dur...
متن کاملCorrecting for Strong Eddy Current Induced B0 Modulation Enables Two-Spoke RF Pulse Design with Parallel Transmission: Demonstration at 9.4T in the Human Brain
Successful implementation of homogeneous slice-selective RF excitation in the human brain at 9.4T using 16-channel parallel transmission (pTX) is demonstrated. A novel three-step pulse design method incorporating fast real-time measurement of eddy current induced B0 variations as well as correction of resulting phase errors during excitation is described. To demonstrate the utility of the propo...
متن کامل