Evaluation of Sanofi Diagnostics Pasteur Chlamydia Microplate EIA shortened assay and comparison with cell culture and Syva Chlamydia MicroTrak II EIA in high- and low-risk populations.
نویسندگان
چکیده
Seven hundred thirty-two female urogenital samples were collected for Chlamydia trachomatis testing by both the Sanofi Diagnostics Pasteur (Chaska, Minn.) Chlamydia Microplate EIA by the shortened protocol and the Syva (San Jose, Calif.) MicroTrak II EIA, and the results were compared with those obtained by cell culture. For the analysis of samples from female patients, the patients were divided into high- and low-risk categories. An additional 121 male urethral samples were collected and tested by the Sanofi Microplate EIA and cell culture; for the analysis of samples from male patients, the patients were divided into asymptomatic and symptomatic categories. All specimens positive by enzyme immunoassay (EIA) were confirmed by a blocking assay following the respective manufacturer's instructions. Specimens negative by EIA that fell within a gray zone 30% below the cutoff and negative cultures with one or more corresponding positive EIA results were tested further by cytocentrifugation and direct immunofluorescent assay. The overall sensitivity, specificity, positive predictive value, and negative predictive value for Syva versus culture were 94, 98.8, 85.5 and 99.6%, respectively. After resolution, the results were 94.5, 99.6, 94.5, and 99.6%, respectively. The parallel results for the Sanofi Microplate EIA versus culture were 94.0, 98.7, and 83.9, and 99.6%, respectively, and after being resolved, the results were 94.9, 100, 100, and 99.6%, respectively. In the small male population tested, the resolved results of the Sanofi Microplate EIA versus culture demonstrated sensitivity, specificity, positive predictive value, and negative predictive value of 100, 100, 100, and 100%, respectively. The present study demonstrated that the Sanofi Microplate EIA shortened protocol is highly sensitive and specific in comparison with cell culture and the Syva MicroTrak II EIA.
منابع مشابه
Comparison of the Syva MicroTrak enzyme immunoassay and Gen-Probe PACE 2 with cell culture for diagnosis of cervical Chlamydia trachomatis infection in a high-prevalence female population.
Culture is currently considered the "gold standard" for detecting Chlamydia trachomatis infections. We evaluated the Syva MicroTrak enzyme immunoassay (EIA) and Gen-Probe PACE 2 tests, which detect chlamydial antigens and rRNA, respectively. These assays were compared with each other and with culture for the detection of C. trachomatis in cervical specimens obtained from 217 women attending a c...
متن کاملHead-to-head evaluation of five chlamydia tests relative to a quality-assured culture standard.
Nucleic acid amplification tests offer superior sensitivity for the detection of Chlamydia trachomatis infection, but many laboratories still use nonamplification methods because of the lower cost and ease of use. In spite of their availability for more than a decade, few studies have directly compared the nonamplification tests. Such comparisons are still needed in addition to studies that dir...
متن کاملEvaluation of Syva enzyme immunoassay for detection of Chlamydia trachomatis in genital specimens.
Detection of Chlamydia trachomatis infection was evaluated by culture and a new Syva enzyme immunoassay (EIA) in 1,012 patients at two Baltimore, Md., sexually transmitted disease clinics. The overall chlamydia prevalence determined by culture was 12%. For 506 fresh cervical and urethral specimens, the sensitivity of Syva EIA was 90% and its specificity was 94% compared with culture. Discordant...
متن کاملEffect of patient characteristics on performance of an enzyme immunoassay for detecting cervical Chlamydia trachomatis infection.
We compared the performance of a commercial enzyme immunoassay (EIA) (Chlamydiazyme; Abbott Diagnostics, North Chicago, Ill.) with that of cell culture for the detection of Chlamydia trachomatis cervical infection in 1,417 women attending public health clinics. Confirmatory chlamydial testing by a direct fluorescent-antibody test (MicroTrak; Syva Co., Palo Alto, Calif.) was performed on specime...
متن کاملEvaluation of an enzyme-linked immunoassay and confirmatory test for the detection of Chlamydia trachomatis in male urine samples.
First pass urine (FPU) samples were compared with urethral swab culture from 304 males attending a genitourinary medicine clinic using an enzyme immuno assay (EIA). All of the EIA positive samples were retested by incorporating a novel blocking reagent into the EIA protocol; 101 were positive by culture of which 83 were also positive by FPU EIA, an additional four were detected in FPU only and ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical microbiology
دوره 33 11 شماره
صفحات -
تاریخ انتشار 1995