The impact of self-shielded cyclotron operation on small-animal PET/CT equipment installed nearby, on the floor just above.
نویسندگان
چکیده
PURPOSE The purpose of this study was to evaluate the impact of a cyclotron on small-animal PET equipment installed directly above the cyclotron. METHODS The cyclotron equipment was HM-12, which has two targets, and the PET/CT equipment was Inveon. The equipment was installed in conformity to Japanese law and regulations. Before installation of the PET/CT equipment, the radiation dose, radio waves, and static and fluctuating magnetic fields were measured at the position where it would be placed, both when the cyclotron was in use and when it was not in use. After installation of the PET/CT, natural background and emission counts were measured at the same place under the same conditions. RESULTS An increase of radiation dose was observed when the target nearest the PET equipment was used. There were no distinct effects of radio waves or static and fluctuating magnetic fields. A significant increase of emission counts, approximately 300 cpm, was observed when the nearest target was used. CONCLUSIONS Though radio waves and static and fluctuating magnetic fields generated by running cyclotron had no influence, a significant increase in emission count was observed. Careful attention should be paid to this influence when very low-radioactivity PET measurements are done.
منابع مشابه
Personalized 18FDG Dose Synthesis Using BG-75 Generator: 1st Year Experience at JCI Accredited Tertiary Care Hospital in Pakistan
Background: Compact cyclotrons are getting popular to fulfill enormous current demands of PET tracers. Aga Khan University Hospital, Karachi, Pakistan has acquired the first smallest footprint of BG-75 Generator for 18FDG-based PET/CT clinical imaging. We are sharing our experience of BG-75 in the first year (December 2015-November 2016) after commissioning. Mater...
متن کاملComparison of PET/CT and CT-based tumor delineation and its effects on the radiation treatment planning for non-small cell lung cancer
Introduction: Tumor volume delineation is the most important step in the radiation treatment planning. In this study the impact of PET/CT data on the tumor delineation precision of non-small cell lung cancer (NSCLC) was investigated. Methods: PET/CT images of 20 patients with primary NSCLC were obtained and imported to the treatment planning system for im...
متن کاملC-11 Choline and FDG PET/CT Imaging of Primary Cholangiocarcinoma – a Comparative Analysis
Objective(s): This study aimed to compare the diagnostic values of 11C-choline and 18F-fluorodeoxyglucose (18F-FDG) positron emission tomography/computed tomography (PET/CT) in patients with cholangiocarcinoma (CCA). Methods: This prospective study was conducted on 10 patients (6 males and 4 females), aged 42-69 years, suspected of having CCA based on CT or magnetic resonance imaging (MRI) resu...
متن کاملEvaluation of Radiation Exposure to Staff and Environment Dose from [18F]-FDG in PET/CT and Cyclotron Center using Thermoluminescent Dosimetry
Background: PET/CT imaging using [18F]-FDG is utilized in clinical oncology for tumor detecting, staging and responding to therapy procedures. Essential consideration must be taken for radiation staff due to high gamma radiation in PET/CT and cyclotron center. The aim of this study was to assess the staff exposure regarding whole body and organ dose and to evaluate environment dose in PET/CT an...
متن کاملTriage of Limited Versus Extensive Disease on 18F-FDG PET/CT Scan in Small Cell lung Cancer
Objective(s): Small cell lung cancer (SCLC) is an aggressive neuroendocrine carcinoma, which accounts for 10-15% of pulmonary cancers and exhibits early metastatic spread. This study aimed to determine the added value of 18F-FDG PET/CT imaging in tumor, node, and metastasis (TNM) staging of SCLC, compared to the conventional computed tomography (CT) scan and its potential role as a prognosticat...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The journal of medical investigation : JMI
دوره 61 1-2 شماره
صفحات -
تاریخ انتشار 2014