In vitro assessment of torque and force generated by novel ProTaper Next Instruments during simulated canal preparation.
نویسندگان
چکیده
INTRODUCTION The purpose of this study was to assess torque and force for simulated canal preparation with a new root canal instrument, ProTaper Next. METHODS Six sets of ProTaper Next Instruments (X1-X5) were used to prepare 36 artificial canals. Files were divided into 6 groups. Different settings of rotations per minute (250, 300, and 350 rpm) and numbers of in-and-out movements to reach working length (3 or 4 insertions [ins]) were applied in each group (250 rpm/3 ins, 250 rpm/4 ins, 300 rpm/3 ins, 300 rpm/4 ins, 350 rpm/3 ins, and 350 rpm/4 ins) by using an automated torque bench. Peak torques (Ncm) as well as positive and negative forces (N) were registered. Analysis of variance and Tukey post hoc tests were applied. Preliminary data for angle and stationary torque at failure were also obtained and compared with peak torque for each instrument. RESULTS Significant differences in peak torque (P < .0001), positive force (P < .002), and negative force (P < .0001) were found for ProTaper Next instruments overall. X2 showed the highest torque with all settings. X5 showed the highest positive force in all groups. X1 and X2 showed the highest negative forces for all groups except for 350 rpm/4 ins. Significantly lower torque (P < .0001) and positive force (P < .007) were measured in the group 350 rpm/4 ins for all instruments except for X4. In contrast, X1 showed a significantly lower negative force for 350 rpm/4 ins. Torque at failure according to American Dental Association no. 28/ISO 36030-1 was lower for X1, X2, and X3 than torque during simulated canal preparation (P < .0001). CONCLUSIONS Under the conditions of this study, using ProTaper Next at 350 rpm and with 4 in-and-out movements resulted in lowest levels of peak torque as well as positive and negative forces.
منابع مشابه
ProTaper rotary root canal preparation: assessment of torque and force in relation to canal anatomy.
AIM To investigate physical parameters for ProTaper nickel-titanium (NiTi) rotary instruments whilst preparing curved canals in maxillary molars in vitro. METHODOLOGY A novel torque-testing platform was used to prepare root canals in 15 extracted human maxillary molars with ProTaper rotary instruments. Peak torque and force was registered along with numbers of rotations required to shape the ...
متن کاملAn in vitro assessment of the physical properties of novel Hyflex nickel-titanium rotary instruments.
AIM To determine several properties including torsional and fatigue limits, as well as torque during canal preparation, of Hyflex, a rotary instrument manufactured from so-called controlled memory nickel-titanium alloy. METHODOLOGY The instruments were tested in vitro using a special torque bench that permits both stationary torque tests according to ISO3630-1 and fatigue limit determination,...
متن کاملIn-vitro Comparison of NiTi Tee, RaCe and Protaper Instruments in Dentinal Crack Formation
Background and Aim:Root canal preparation with rotary instruments may cause dentinal cracks leading to tooth fracture. The aim of this study was to compare three different rotary systems ProTaper, RaCe and Niti Tee on formation of dentinal cracks following root canal preparation. Materials and Methods: In this experimental study, 50 extracted mandibular first molars were selected. Teeth having ...
متن کاملمقایسه تأثیر آمادهسازی کانال دندان به روش استپ بک با استفاده از فایلهای دستی و گیتس گلیدن، با سیستم روتاری پروتیپر در مقاومت ریشه به شکستگی عمودی
Introduction: Cleaning and shaping of the root canal system with an efficient and safe technique are the major goals of root canal treatment. The aim of this study was to compare the conventional root canal preparation technique by hand instruments and Gates Glidden drills with ProTaper Universal Rotary system on the root susceptibility to vertical fracture. Methods: Thirty extracted human man...
متن کاملStereomicroscopic Evaluation of Radicular Dentinal Defects after Using EDTA Gel with Different Rotary Systems: An In-Vitro Study
The purpose of the present study was to study the occurrence of radicular dentinal defects after using EDTA gel during preparation of the root canal with ProTaper Universal (Dentsply Maillefer, Switzerland), HyFlex (ColteneWhaledent, Allstetten, Switzerland), and ProTaper Next (Dentsply Maillefer) rotary instruments. Forty-eight single rooted premolars were selected. Twelve teeth were instrumen...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of endodontics
دوره 39 12 شماره
صفحات -
تاریخ انتشار 2013