Digital pulse-shape discrimination of fast neutrons and γ rays
نویسنده
چکیده
Discrimination of the detection of fast neutrons and γ rays in a liquid scintillator detector has been investigated using digital pulse-processing techniques. An experimental setup with a 252Cf source, a BC-501 liquid scintillator detector, and a BaF2 detector was used to collect waveforms with a 100 Ms/s, 14 bit sampling ADC. Three identical ADC’s were combined to increase the sampling frequency to 300 Ms/s. Four different digital pulse-shape analysis algorithms were developed and compared to each other and to data obtained with an analogue neutron-γ discrimination unit. Two of the digital algorithms were based on the charge comparison method, while the analogue unit and the other two digital algorithms were based on the zero-crossover method. Two different figure-of-merit parameters, which quantify the neutron-γ discrimination properties, were evaluated for all four digital algorithms and for the analogue data set. All of the digital algorithms gave similar or better figure-of-merit values than what was obtained with the analogue setup. A detailed study of the discrimination properties as a function of sampling frequency and bit resolution of the ADC was performed. It was shown that a sampling ADC with a bit resolution of 12 bits and a sampling frequency of 100 Ms/s is adequate for achieving an optimal neutron-γ discrimination for pulses having a dynamic range for deposited neutron energies of 0.3-12 MeV. An investigation of the influence of the sampling frequency on the time resolution was made. A FWHM of 1.7 ns was obtained at 100 Ms/s.
منابع مشابه
Neutron generator burst timing measured using a pulse shape discrimination plastic scintillator with silicon photomultiplier readout
An EJ-299-34 plastic scintillator with silicon photomultiplier (SiPM) readout was used to measure the fast neutron output of a pulsed Thermo-Fisher A-325 Deuterium-Tritium sealed tube neutron generator (STNG). The SiPM signals were handled by a prototype digital pulse processing system, based on a free-running analogue to digital converter feeding a digital signal processor (DSP). Pulse shape d...
متن کاملDigital pulse shape analysis for the capture-gated liquid scintillator BC-523A
The BC-523A detector is a boron-10-loaded liquid scintillation detector that is sensitive to not only fast neutrons and gamma rays, but also to thermal neutrons. This detector has recently attracted the attention of several researchers and has shown promise for neutron spectroscopy. Several efforts have been made to accurately characterize this detector for several neutron and gamma-ray energie...
متن کامل“Neutron Shell”: A high efficiency array of neutron detectors for γ−ray spectroscopic studies with Gammasphere
A shell of neutron detectors was designed, constructed, and employed in γ−ray spectroscopy with Gammasphere. It consists of up to 35 tapered regular hexagons that replace the same number of forward Ge-detector modules in Gammasphere. The shell was designed for high detection efficiency and very good neutron−γ discrimination. The simultaneous use of time of flight, and two methods of pulse shape...
متن کاملPulse-shape Discrimination for Identification of Neutron Sources Using the Bc-501a Liquid Scintillator
Fast and accurate identification of neutron sources is very important in nuclear nonproliferation and international safeguards applications. Commonly, the identification of neutron spectra relies on unfolding procedures, which requires the solution of an ill-posed problem. In this paper, we present a new application of an existing technique, which can be used for neutron source identification w...
متن کاملDigital discrimination of neutrons and gamma-rays in liquid scintillators using wavelets
A novel algorithm for the digital discrimination of neutrons and gamma-rays in a mixed radiation field is presented. Most of the pulse shape discrimination methods in scintillation detection systems use time-domain features of the signal (e.g. charge comparison method or constant-time discrimination). However, there are no frequency-domain discrimination methods up to date in the literature. Ou...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -