Feasibility of CBCT‐based dose with a patient‐specific stepwise HU‐to‐density curve to determine time of replanning
نویسندگان
چکیده
PURPOSE (a) To investigate the accuracy of cone-beam computed tomography (CBCT)-derived dose distributions relative to fanbeam-based simulation CT-derived dose distributions; and (b) to study the feasibility of CBCT dosimetry for guiding the appropriateness of replanning. METHODS AND MATERIALS Image data corresponding to 40 patients (10 head and neck [HN], 10 lung, 10 pancreas, 10 pelvis) who underwent radiation therapy were randomly selected. Each patient had both intensity-modulated radiation therapy and volumetric-modulated arc therapy plans; these 80 plans were subsequently recomputed on the CBCT images using a patient-specific stepwise curve (Hounsfield units-to-density). Planning target volumes (PTVs; D98%, D95%, D2%), mean dose, and V95% were compared between simulation-CT-derived treatment plans and CBCT-based plans. Gamma analyses were performed using criterion of 3%/3 mm for three dose zones (>90%, 70%~90%, and 30%~70% of maximum dose). CBCT-derived doses were then used to evaluate the appropriateness of replanning decisions in 12 additional HN patients whose plans were previously revised during radiation therapy because of anatomic changes; replanning in these cases was guided by the conventional observed source-to-skin-distance change-derived approach. RESULTS For all disease sites, the difference in PTV mean dose was 0.1% ± 1.1%, D2% was 0.7% ± 0.1%, D95% was 0.2% ± 1.1%, D98% was 0.2% ± 1.0%, and V95% was 0.3% ± 0.8%; For 3D dose comparison, 99.0% ± 1.9%, 97.6% ± 4.4%, and 95.3% ± 6.0% of points passed the 3%/3 mm criterion of gamma analysis in high-, medium-, and low-dose zones, respectively. The CBCT images achieved comparable dose distributions. In the 12 previously replanned 12 HN patients, CBCT-based dose predicted well changes in PTV D2% (Pearson linear correlation coefficient = 0.93; P < 0.001). If 3% of change is used as the replanning criteria, 7/12 patients could avoid replanning. CONCLUSIONS CBCT-based dose calculations produced accuracy comparable to that of simulation CT. CBCT-based dosimetry can guide the decision to replan during the course of treatment.
منابع مشابه
Evaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners
Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsf...
متن کاملA feasibility study on the use of MV-CBCT images for urgent palliative treatment planning
Introduction: The application of 3D volumetric imaging modalities in treatment planning of radiation therapy can provide more precisely define tumor localization, and computed tomography (CT) is the most common accepted method for treatment planning. Given the lack of a CT scanner stationed in all radiotherapy departments and equipping most of the medical linear accelerators wi...
متن کاملStudy of Variation in Dose Calculation Accuracy Between kV Cone-Beam Computed Tomography and kV fan-Beam Computed Tomography
Cone-beam computed tomography (CBCT) images are presently used for geometric verification for daily patient positioning. In this work, we have compared the images of CBCT with the images of conventional fan beam CT (FBCT) in terms of image quality and Hounsfield units (HUs). We also compared the dose calculated using CBCT with that of FBCT. Homogenous RW3 plates and Catphan phantom were scanned...
متن کاملInvestigation of the usability of conebeam CT data sets for dose calculation
BACKGROUND To investigate the feasibility and accuracy of dose calculation in cone beam CT (CBCT) data sets. METHODS Kilovoltage CBCT images were acquired with the Elekta XVI system, CT studies generated with a conventional multi-slice CT scanner (Siemens Somatom Sensation Open) served as reference images. Material specific volumes of interest (VOI) were defined for commercial CT Phantoms (CA...
متن کاملEvaluation of Effect of Different Computed Tomography Scanning Protocols on Hounsfield Unit and Its Impact on Dose Calculation by Treatment Planning System
Introduction: In radiotherapy treatment planning system (TPS), basic input is the data from computed tomography (CT) scan, which takes into account the effect of inhomogeneities in dose calculations. Measurement of CT numbers may be affected by scanner-specific parameters. Therefore, it is important to verify the effect of different CT scanning protocols on Hounsfield unit (HU) and its impact o...