The goal is to help create smooth energy-optimal monophasic pulse waveforms for defibrillation using the Luo-Rudy cardiomyocyte membrane computer model. The waveforms were described with the help of the piecewise linear function. Each line segment provides a transition from one present level of the transmembrane potential to the next with a minimal energy value. The duration of the last segment was defined as a minimum duration at which an action potential occurs. Monophasic waveforms of segments 3, 10 and 29 were built using different increments of the transmembrane potential. The pulse energy efficiency was evaluated according to their threshold energy ratios in mA2∙ms/cm4. There was virtually no difference between the threshold energy ratios of the three waveforms constructed and those of the previously studied energy-optimal halfsine waveform: 241 242 and 243 mA2∙ms/cm4. The pulse waveform constructed is characterized by a low rise and fall as the duration of the rise is ~1.5 times longer than that of the fall. Conclusion: Energy-optimal smooth monophasic pulse waveforms have the same threshold energy ratio as the optimal half-sine one which was studied before. The latter is equivalent to the first phase of biphasic quasi-sinusoidal Gurvich-Venin pulse which has been used in Russia since 1972. Thus, the use of the Luo-Rudy cardiomyocyte membrane model appears to offer no possibilities for a substantial increase in the energy efficiency (threshold energy ratio reduction) of the classical monophasic defibrillation pulse waveforms. Corresponding author. V. A. Vostrikov et al.
منابع مشابه
Fully discharging phases. A new approach to biphasic waveforms for external defibrillation.
BACKGROUND Phase-2 voltage and maximum pulse width are dependent on phase-1 pulse characteristics in a single-capacitor biphasic waveform. The use of 2 separate output capacitors avoids these limitations and may allow waveforms with lower defibrillation thresholds. A previous report also suggested that the optimal tilt may be >70%. This study was designed to determine an optimal biphasic wavefo...
متن کاملOptimization of biphasic waveforms for human nonthoracotomy defibrillation.
BACKGROUND Biphasic waveforms reduce defibrillation threshold (DFT) in a wide variety of models. Although there are several human studies of long-duration, high-tilt biphasic waveform defibrillation, the specific biphasic waveform shape required to achieve optimal DFT reduction is unknown. METHODS AND RESULTS This study tested the effect of single capacitor biphasic waveform tilt modification...
متن کاملStrength-duration and probability of success curves for defibrillation with biphasic waveforms.
Certain biphasic waveforms require less energy to defibrillate than do monophasic pulses of equal duration, although the mechanisms of this increased effectiveness remain unclear. This study used strength-duration and percent success curves for defibrillation with monophasic and biphasic truncated exponential waveforms to explore these mechanisms. In part 1, defibrillation thresholds were deter...
متن کاملOptimal small-capacitor biphasic waveform for external defibrillation: influence of phase-1 tilt and phase-2 voltage.
BACKGROUND Biphasic waveforms have been reported to be more efficacious than monophasic waveforms for external defibrillation. This study examined the optimal phase-1 tilts and phase-2 leading-edge voltages with small capacitors (60 and 20 microF) for external defibrillation. We also assessed the ability of the "charge-burping" model to predict the optimal waveforms. METHODS AND RESULTS Two g...
متن کاملA prospective randomized evaluation of biphasic versus monophasic waveform pulses on defibrillation efficacy in humans.
Biphasic waveforms have been suggested as a superior waveform for ventricular defibrillation. To test this premise, a prospective randomized intraoperative evaluation of defibrillation efficacy of monophasic and biphasic waveform pulses was performed in 22 survivors of out of hospital ventricular fibrillation who were undergoing cardiac surgery for implantation of an automatic defibrillator. Th...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2015