Percent depth‐dose distribution discrepancies from very small volume ion chambers

نویسندگان

  • Vikren Sarkar
  • Brian Wang
  • Hui Zhao
  • Bart Lynch
  • Joshua A. James
  • Kiernan T. McCullough
  • Bill J. Salter
چکیده

As very small ion chambers become commercially available, medical physicists may be inclined to use them during the linear accelerator commissioning process to better characterize the beam in steep dose gradient areas. For this work, a total of eight different ion chambers (volumes from 0.007 cc to 0.6 cc) and four different scanning systems were used to scan PDDs at both +300V and -300V biases. We observed a reproducible, significant difference (overresponse with depth) in PDDs acquired when using very small ion chambers, with specific bias/water tank combinations - up to 5% at a depth of 25 cm in water. This difference was not observed when the PDDs were sampled using the ion chamber in static positions in conjunction with an external electrometer. This suggests noise/signal interference produced by the controller box and cable system assemblies, which can become relatively significant for the very small current signals collected by very small ion chambers, especially at depth as the signal level is even further reduced. Based on the results observed here, the use of very small active volume chambers under specific scanning conditions may lead to collection of erroneous data, introducing systematic errors into the treatment planning system. In case the use of such a chamber is required, we recommend determining whether such erroneous effect exists by comparing the scans with those obtained with a larger chamber.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Introduction For dose measurement in Megavoltage (MV) photon beams with ion chambers, the effect of volume occupied by the air cavity is not negligible. Therefore, the result of measurement should be corrected with a displacement perturbation correction factor (Pdis) or using an effective point of measurement (EPOM). The aim of this study is to calculate the EPOM for cylindrical ion chamber and...

متن کامل

Dosimetric Evaluation of Linac Photon Small Fields using MAGIC Polymer Gels

Introduction: In radiotherapy, methods of treatment planning are becoming increasingly more complicated. This requires verification of the doses delivered to increasingly smaller and more precise regions. Radiotherapy techniques are continuously employing smaller and smaller field sizes to deliver tighter radiation doses with higher therapeutic ratios, generating interest among researchers to p...

متن کامل

Small Volume Ionization Chambers Angular Dependence and Its Influence on Point-Dose Measurements

Purpose: Cylindrical ionization chambers as the PinPoint and the CC01 have high-spatial resolution and cylindrical symmetry in relevance to small fields. Dosimetry of newly advanced technologies that include non-coplanar beams, such as in Gamma-Knife, VMAT and IMRT treatment plans, the angular dependence of the ion chamber is essential. Therefore in this study we simulated two common used ion c...

متن کامل

Comment on "Comparison of ionization chambers of various volumes for IMRT absolute dose verification".

IMRT plans are usually verified by phantom measurements: dose distributions are measured using film and the absolute dose using an ionization chamber. The measured and calculated doses are compared and planned MUs are modified if necessary. To achieve a conformal dose distribution, IMRT fields are composed of small subfields, or "beamlets." The size of beamlets is on the order of 1 x 1 cm2. The...

متن کامل

Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams

The International Atomic Energy Agency Technical Reports Series-398 code of practice for dosimetry recommends measuring photon percentage depth dose (PDD) curves with parallel-plate chambers. This code of practice was published before flattening filter-free (FFF) beams were widely used in clinical linear accelerators. The choice of detector for PDD measurements needs to be reassessed for FFF be...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015