A Comparative Study of Blood Glucose Measurements Using Glucometer Readings and the Standard Method in the Diagnosis of Neonatal Hypoglycemia
نویسندگان
چکیده
Background: Hypoglycemia is one of the most common neonatal disorders, associated with severe complications. There has been a great deal of controversy regarding the definition and screening of hypoglycemia. Therefore, in this study, we aimed to determine a cut-off value for blood glucose level in glucometer readings. Methods: This cross-sectional study was conducted on 238 newborns at risk of hypoglycemia, admitted to Baqiyatallah Hospital of Tehran, Iran in 2012; the subjects were selected via simple sampling. After obtaining informed consents from the newborns’ parents, 1 cc blood samples were sent to the laboratory for measuring the blood glucose level. Moreover, venous blood samples, as well as heel-stick blood samples, were obtained for glucometer measurements. Blood glucose measurements were used to determine the cut-off value by the receiver operating characteristic (ROC) curve and make comparisons with the diagnostic criteria for hypoglycemia in the literature. Results: A total of 238 infants with the mean weight of 2869±821.9 g were enrolled in this study. The mean (±SD) blood glucose levels were 65.1±22.9, 82.9±24.7, and 84.4±24.8 mg/dl, based on the standard laboratory method, glucometer reading of venous blood samples, and glucometer reading of heel-stick capillary blood samples, respectively. The optimal cut-off point for hypoglycemia was determined as 65 mg/dl, using glucometer-based assessment of heel-stick blood samples. Conclusion: The significant difference in blood glucose levels measured by the laboratory method and outpatient glucometer readings highlights the importance of a cut-off value for rapid assessment and control of blood glucose and timely detection of hypoglycemia. In fact, the cut-off value introduced in the present study could facilitate such measurements.
منابع مشابه
A Comparative Study of Blood Glucose Measurements Using Glucometer Readings and the Standard Method in the Diagnosis of Neonatal Hypoglycemia
Background: Hypoglycemia is one of the most common neonatal disorders, associated with severe complications. There has been a great deal of controversy regarding the definition and screening of hypoglycemia. Therefore, in this study, we aimed to determine a cut-off value for blood glucose level in glucometer readings. Methods: This cross-sectional study was conducted on 238 newborns at risk of ...
متن کاملبررسی ارزش گلوکومتری در تشخیص هیپوگلیسمی نوزادان
Background and purpose: Hypoglycemia is an important problem in newborns which requires early diagnosis and treatment. Laboratory testing of blood glucose level and glucometer readings are used to detect hypoglycemia but sometimes the results are given late. This study aimed at evaluating the accuracy of glucometer in detection of hypoglycemia in newborns at neonate intensive care unit (NICU), ...
متن کاملمقایسه میزان همخوانی یافتههای سنجش قندخون با استفاده از گلوکومتر ON CALL و گلوکوکارد با روش آزمایشگاهی استاندارد
Background: Diabetes Mellitus (DM) is one the most common chronic disease, with many complications including renal failure, blindness and non traumatic amputation. Prevention of complications is possible through monitoring and control of blood glucose levels. Considering how easy home blood glucose monitoring is, we decided to evaluate the performance of two available glucometers for detection...
متن کاملBlood glucose measurement by glucometer in comparison with standard method in diagnosis of neonatal hypoglycemia.
Hypoglycemia is considered as a serious risk factor in neonates. In the majority of cases, it occurs with no clinical symptoms. Accordingly, early diagnosis is extremely imperative, which can also lead to less morbidity and mortality. The aim of this study was to assess the importance of screening blood glucose using glucometer (known as a quick and cost-effective diagnostic test) in comparison...
متن کاملمقایسه قدرت دستگاه گلوکومتر Accu-chek با Elite در تشخیص دیابت با استفاده از خونریزی متعاقب Probing
Abstract Background: Recent studies have assessed the validity of glucometer devices in estimating the glucose level of gingival blood during routine periodontal examination (Bleeding On Probing). The purpose of this study was to evaluate whether blood oozing from gingival tissues during routine periodontal examination can be used for determining g...