Gum elastic bougie-guided insertion of the ProSeal Laryngeal Mask Airway.

نویسندگان

  • J Brimacombe
  • C Keller
چکیده

We tested the hypothesis that gum elastic-bougie-guided insertion of the ProSeal Laryngeal Mask Airway is more frequently successful than introducer tool guided insertion after failed digital insertion. One hundred anaesthetized patients (ASA 1-2, aged 18 to 80 years) were randomized for the second insertion attempt using either the gum elastic bougie-guided or introducer tool techniques. The bougie-guided technique involved priming the drain tube with the bougie, placing the bougie in the oesophagus using laryngoscope guidance, digital insertion along the palato-pharyngeal curve, and bougie removal. The introducer tool technique involved attaching the introducer tool, single-handed rotation along the palatopharyngeal curve, and introducer tool removal. Failed insertion was classified as (i) failed passage into the pharynx, (ii) malposition, or (iii) ineffective ventilation. Any blood staining was documented. Insertion was more frequently successful (50/50 vs 15/50, P=0.0002) and faster (35+/-17 s vs 54+/-45 s, mean+/-SD, P=0.006) with the bougie-guided technique. All failed insertions with the introducer tool technique were successful with the bougie-guided technique. The aetiology of failed insertion was similar for the digital and introducer tool techniques in 94% (33/35) of patients. There was no blood staining on the bougie, laryngoscope or introducer tool at removal, but blood staining was more common on the ProSeal Laryngeal Mask Airway with the introducer tool technique (9/50 vs 2/50, P=0.03). We conclude that the gum elastic bougie-guided insertion has a higher success rate and causes less trauma than the insertion tool insertion technique after failed digital insertion of the ProSeal Laryngeal Mask Airway.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Successful delivery in a morbidly obese patient after failed intubation and regional technique.

a review of the literature. Can J Anaesth 2005; 52: 739–60 7 Brimacombe J, Keller C, Judd DV. Gum elastic bougie-guided insertion of the ProSeal laryngeal mask airway is superior to the digital and introducer tool techniques. Anesthesiology 2004; 100: 25–9 8 Craven RM, Laver SR, Cook TM, Nolan JP. Use of the Pro-Seal LMA facilitates percutaneous dilatational tracheostomy. Can J Anaesth 2003; 50...

متن کامل

Comparative evaluation of gum-elastic bougie and introducer tool as aids in positioning of ProSeal laryngeal mask airway in patients with simulated restricted neck mobility

BACKGROUND The ProSeal laryngeal mask airway (PLMA) is a unique laryngeal mask with a modified cuff to improve seal and a channel to facilitate gastric tube placement. This is a better device in difficult airway situations compared to classic laryngeal mask airway. This prompted us to study the ease of insertion and positioning of PLMA in patients with simulated restricted neck mobility while u...

متن کامل

An unusual cause of airway obstruction after guided insertion of the ProSeal LMA.

The ProSealTM laryngeal mask airway (PLMA) insertion using a gum-elastic bougie placed in the oesophagus as a guide has a high first-attempt success rate ( 99%) and a potential role in difficult airway management [1]. On those occasions when it fails, the aetiology is usually laryngospasm, severe epiglottic downfolding, glottic/supraglottic compression or cuff infolding [2]. Most of these probl...

متن کامل

Management of the anticipated difficult airway.

The protocol dictated the use of a gum elastic bougie (GEB) for intubation difficulty, followed by insertion of an intubating laryngeal mask airway (ILMA) if intubation was still not successful. Intubation difficulty was defined as two failed laryngoscopy attempts by a senior physician using optimal laryngeal manipulation. For difficult ventilation, the protocol recommended ILMA insertion follo...

متن کامل

Ease of Proseal Laryngeal Mask Airway Insertion Using Miller Blade; a Comparison with Conventional Digital Technique

The Proseal laryngeal mask airway (PLMA) (The Laryngeal Mask Company Ltd., Victoria, Seychelles) is an advanced form of the LMA with a softer and larger cuff to improve sealing and a drain tube to permit venting of the stomach. However, when a larger cuff is used, it is more difficult to save space for inserting the PLMA in the patient's mouth, and the larger cuff is more likely to fold over (1...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Anesthesiology

دوره 100 1  شماره 

صفحات  -

تاریخ انتشار 2004