Low-dose computed tomography of the lumbar spine: a phantom study on imaging parameters and image quality.

نویسندگان

  • Muhammed Alshamari
  • Mats Geijer
  • Eva Norrman
  • Håkan Geijer
چکیده

BACKGROUND Lumbar spine radiography has limited diagnostic value but low radiation dose compared with computed tomography (CT). The average effective radiation dose from lumbar spine radiography is about 1.1 mSv. Low-dose lumbar spine CT may be an alternative to increase the diagnostic value at low radiation dose, around 1 mSv. PURPOSE To determine the optimal settings for low-dose lumbar spine CT simultaneously aiming for the highest diagnostic image quality possible. MATERIAL AND METHODS An ovine lower thoracic and lumbar spine phantom, with all soft tissues around the vertebrae preserved except the skin, was placed in a 20 L plastic container filled with water. The phantom was scanned repeatedly with various technical settings; different tube potential, reference mAs, and with different convolution filters. Five radiologists evaluated the image quality according to a modification of the European guidelines for multislice computed tomography (MSCT) quality criteria for lumbar spine CT 2004. In a visual comparison the different scans were also ranked subjectively according to perceived image quality. Image noise and contrast were measured. RESULTS A tube potential of 120 kV with reference mAs 30 and medium or medium smooth convolution filter gave the best image quality at a sub-millisievert dose level, i.e. with an effective dose comparable to that from lumbar spine radiography. CONCLUSION Low-dose lumbar spine CT thus opens a possibility to substitute lumbar spine radiography with CT without obvious increase in radiation dose.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography

Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedures to reduce radiation dose without compromising the image quality. Material and Methods: This ex...

متن کامل

Low-dose computed tomography of the abdomen and lumbar spine

Muhammed Alshamari (2016): Low-dose computed tomography of the abdomen and lumbar spine. Örebro Studies in Medicine 137. Radiography is a common radiologic investigation despite abundant evidence of its limited diagnostic value. On the other hand, computed tomography (CT) has a high diagnostic value and is widely considered to be among the most important advances in medicine. However, CT expose...

متن کامل

Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography

Introduction: Digital radiography possesses a wide dynamic range and has a major advantage in producing an acceptable image of diagnostic value even though overexposure occurs. Lumbar spine (LS) radiography is the most common examinations that gives high radiation dose to patients and accounts for the highest collective population dose of any conventional radiographic ...

متن کامل

Assessing the image quality and eye lens dose reduction using bismuth shielding in computed tomography of brain

Background: Epidemiological studies show that computed tomography (CT) is one of the main sources of ionizing radiations. Shielding of radiosensitive organs is one of the dose reduction methods. This study aimed to assess the eye lens dose reduction and image quality resulting from the use of radio-protective bismuth shield in brain CT imaging. Methods: Bismut...

متن کامل

Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality

Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. Material and Methods: A poly(me...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Acta radiologica

دوره 55 7  شماره 

صفحات  -

تاریخ انتشار 2014