Oxygen outflow delivered by manually operated self-inflating resuscitation bags in patients breathing spontaneously.
نویسندگان
چکیده
OBJECTIVE To determine the oxygen outflow delivered by seven different models of manually operated self-inflating resuscitation bags (with and without an oxygen reservoir connected), which were tested using different oxygen supply rates without manipulating the bag, by simulating their use in patients breathing spontaneously. METHODS The oxygen outflow was measured using a wall oxygen flow meter and a flow meter/respirometer attached to the bag, together with another flow meter/respirometer attached to the patient connection port. The resuscitation bags that allow the connection of an oxygen reservoir were tested with and without this device. All resuscitation bags were tested using oxygen supply rates of 1, 5, 10, and 15 L/min. Statistical analyses were performed using analysis of variance and t-tests. RESULTS The resuscitation bags that allow the connection of an oxygen reservoir presented a greater oxygen outflow when this device was connected. All resuscitation bags delivered a greater oxygen outflow when receiving oxygen at a rate of 15 L/min. However, not all models delivered a sufficient oxygen outflow even when the two previous conditions were satisfied. CONCLUSIONS Of the resuscitation bags studied, those that allow the connection of an oxygen reservoir must have this reservoir connected to the bag when used as a source of oxygen in nonintubated spontaneously breathing patients. All of the models studied should receive oxygen at a rate > 15 L/min. It is not safe to use manually operated self-inflating resuscitation bags for this purpose without knowing their characteristics.
منابع مشابه
Alterations in peak inspiratory pressure and tidal volume delivered by manually operated self-inflating resuscitation bags as a function of the oxygen supply rate.
OBJECTIVE To assess possible alterations in the tidal volume and peak inspiratory pressure delivered by seven models of manually operated self-inflating resuscitation bags as a function of the oxygen supply rate used. METHODS The resuscitation bags tested were the following: Oxigel, models A and B; CE Reanimadores; ProtecSolutions; Missouri; Axmed; and Narcosul. For the measurements, a wall o...
متن کاملThe self-inflating resuscitation bag delivers high oxygen concentrations when used without a reservoir: implications for neonatal resuscitation.
OBJECTIVE To measure the delivered fractional oxygen concentration (F(DO(2))) from preterm-size Laerdal silicone resuscitators (PLSR) without a reservoir. BACKGROUND The North American Neonatal Resuscitation Program manual states that self-inflating bags without a reservoir deliver approximately 40% oxygen, differing from the PLSR manufacturer's specifications. METHODS A neonatal test lung ...
متن کاملNeonatal and pediatric manual hyperinflation: influence of oxygen flow on ventilation parameters.
BACKGROUND Although self-inflating bags are widely used for manual hyperinflation, they do not allow ventilation parameters, such as pressure or volume, to be set. We studied the ventilation performance of neonatal and pediatric self-inflating bags. METHODS We asked 22 physiotherapists to manually hyperinflate 2 lung models (neonatal and pediatric), using self-inflating bags from 3 manufactur...
متن کاملInnovating Neonatal Resuscitation Bags for Ease of Use in Diverse Locations.
The Chinese emperor Huang-Ti (2698–2599 BCE) noted that newborn death from ventilatory failure was common in premature infants.1 Galen (129–199 CE) inflated the lungs of deceased animals with a bellows and concluded that the air movement causes the chest to rise.1,2 Between 1850 and 1950, many techniques were used to resuscitate newborns, including swinging the infant upside down,1-3 rhythmic t...
متن کاملSmaller self-inflating bags produce greater guideline consistent ventilation in simulated cardiopulmonary resuscitation
BACKGROUND Suboptimal bag ventilation in cardiopulmonary resuscitation (CPR) has demonstrated detrimental physiological outcomes for cardiac arrest patients. In light of recent guideline changes for resuscitation, there is a need to identify the efficacy of bag ventilation by prehospital care providers. The objective of this study was to evaluate bag ventilation in relation to operator ability ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Jornal brasileiro de pneumologia : publicacao oficial da Sociedade Brasileira de Pneumologia e Tisilogia
دوره 34 4 شماره
صفحات -
تاریخ انتشار 2008