In vitro comparative optical bench analysis of a spherical and aspheric optic design of the same IOL model

نویسندگان

  • Tamer Tandogan
  • Gerd U Auffarth
  • Chul Y Choi
  • Stephanie Liebing
  • Christian Mayer
  • Ramin Khoramnia
چکیده

BACKGROUND To analyse objective optical properties of the spherical and aspheric design of the same intraocular lens (IOL) model using optical bench analysis. METHODS This study entailed a comparative analysis of 10 spherical C-flex 570 C and 10 aspheric C-flex 970 C IOLs (Rayner Intraocular Lenses Ltd., Hove, UK) of 26 diopters [D] using an optical bench (OptiSpheric, Trioptics, Germany). In all lenses, we evaluated the modulation transfer function (MTF) at 50 lp/mm and 100 lp/mm and the Strehl Ratio using a 3-mm (photopic) and 4.5-mm (mesopic) aperture. RESULTS At 50 lp/mm, the MTF values were 0.713/0.805 (C-flex 570 C/C-flex 970 C) for a 3-mm aperture and 0.294/0.591 for a 4.5-mm aperture. At 100 lp/mm, the MTF values were 0.524/0.634 for a 3-mm aperture and 0.198/0.344 for a 4.5-mm aperture. The Strehl Ratio was 0.806/0.925 and 0.237/0.479 for a 3-mm and 4.5-mm aperture respectively. A Mann-Whitney U test revealed all intergroup differences to be statistically significant (p < 0.01). CONCLUSION The aspheric IOL design achieved higher MTF values than the spherical design of the same IOL for both apertures. Moreover, the differences between the two designs of the IOL were more prominent for larger apertures. This suggests that the evaluated IOL provides enhanced optical quality to patients with larger pupils or working under mesopic conditions.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Energy balance in apodized diffractive multifocal intraocular lenses

The energy distribution between the distance and near images formed in a model eye by three different apodized diffractive multifocal intraocular lenses (IOLs) is experimentally determined in an optical bench. The model eye has an artificial cornea with positive spherical aberration (SA) similar to human cornea. The level of SA upon the IOL, which is pupil size dependent, is controlled using a ...

متن کامل

Intraocular lens implantation position sensitivity as a function of refractive error.

PURPOSE To quantify the change in ocular aberrations due to intraocular lens (IOL) implantation position errors for pseudo-phakic eyes with different refractive errors. METHODS The theoretical part of this work was based on ray-tracing in emmetropic and myopic eye models. The possible misalignments i.e. the decentration, tilt and axial translation, of the IOLs were statistically combined toge...

متن کامل

Performance of presbyopia-correcting intraocular lenses in distance optical bench tests.

PURPOSE To compare the optical performance of 6 presbyopia-correcting intraocular lenses (IOLs) of different designs (accommodating or multifocal and aspheric or spheric). SETTING Alcon Laboratories, Inc., Fort Worth, Texas, USA. METHODS The optical performance of 6 IOLs were compared using modulation transfer function (MTF) and the United States Air Force 1951 Resolution Target (AFT) testi...

متن کامل

Spherical Aberration of Intraocular Lenses

Advancements in cataract surgery and intraocular lens (IOL) design have optimized the postoperative optical performance of the pseudophakic eye. One of the recent spotlights of IOL design has been formulating optical properties similar to a clear, young lens and addressing spherical aberration. Spherical aberration in the human eye is a combination of the positive spherical aberration of the co...

متن کامل

[Halos and multifocal intraocular lenses: origin and interpretation].

OBJECTIVE To present the theoretical and experimental characterization of the halo in multifocal intraocular lenses (MIOL). METHOD The origin of the halo in a MIOL is the overlaying of 2 or more images. Using geometrical optics, it can be demonstrated that the diameter of each halo depends on the addition of the lens (ΔP), the base power (P(d)), and the diameter of the IOL that contributes to...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2017