Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D

نویسندگان

  • Songlin Yu
  • Weiyan Zhou
  • Xinqi Cheng
  • Huiling Fang
  • Ruiping Zhang
  • Qian Cheng
  • Jianhua Han
  • Wei Su
  • Liangyu Xia
  • Ling Qiu
چکیده

A significant bias was found when using the Siemens ADVIA Centaur XP system for measurement of 25-hydroxyvitamin D (25OHD) with VACUETTE® tubes with Serum Clot Activator and Gel. Here, we examined whether other commonly used tubes or temperatures affected 25OHD results obtained with the Siemens ADVIA Centaur XP system. Serum was collected into five types of vacuum blood collection tubes from three manufacturers, and 25OHD was analyzed using the Siemens ADVIA Centaur XP system and liquid chromatography tandem mass spectrometry (LC-MS/MS) immediately or after storage at 4°C or -80°C for 48 h. Significantly higher 25OHD values were found when using the Siemens ADVIA Centaur XP system with VACUETTE® tubes with serum clot activator and gel and VACUETTE® tubes with clot activator but no gel compared with VACUETTE® tubes with no additives. The 25OHD values in all of these tubes were not significantly different from those obtained by LC-MS/MS. Moreover, after storage at -80°C for 48 h, the values obtained in IMPROVEVACUTER® tubes with serum clot activator and gel significantly increased, with a mean bias of 74.6% compared with the values before storage, on analysis with the Siemens ADVIA Centaur XP system. VACUETTE® tubes containing additives significantly affect the accuracy of 25OHD results obtained using the Siemens ADVIA Centaur XP system. Additionally, the composition of serum collected in IMPROVEVACUTER® tubes was affected by freezing, resulting in different measurements when using the Siemens 25OHD assay platform.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance evaluation of two immunoassays for 25-hydroxyvitamin D

Although immunoassays in measuring 25-hydroxyvitamin D [25(OH)D] have been improved recently, relatively large differences are still seen between results of 25(OH)D measured by immunoassays and by liquid chromatography-tandem mass spectrometry (LC-MS/MS). In the present studies, we compared two immunoassays with LC-MS/MS for measuring 25(OH)D concentrations. Concentrations of 25-hydroxyvitamin ...

متن کامل

Performance characteristics of six automated 25-hydroxyvitamin D assays: Mind your 3s and 2s.

BACKGROUND Measurement of 25(OH)D has evolved rapidly, with an increased number of high-throughput automated immunoassays becoming available in recent years. The aim of this study was to fully evaluate six commercially available automated 25(OH)D immunoassays, including 2 newly available (Beckman Coulter) and one recalibrated (Siemens) assay. Comparisons were made to specifically identify the e...

متن کامل

Influence of the immunoassay used in measurement of serum vitamin D levels.

INTRODUCTION Serum 25-hydroxyvitamin D [25(OH)D] levels are the best indicator of vitamin D levels in the body. Precision, reproducibility, and lack of standardization are the main problems in such measurements. The aim of this study was to compare the 25(OH)D levels measured using Elecsys Vitamin D Total (Roche) and ADVIA Centaur Vitamin D Total (Siemens). MATERIAL AND METHODS 25(OH)D levels...

متن کامل

Performance characteristics of six intact parathyroid hormone assays.

The aim of this study was to evaluate the performance characteristics of 6 intact parathyroid hormone assays: Access 2 (Beckman Coulter, Fullerton, CA), ARCHITECT i2000(SR) (Abbott Diagnostics, Abbott Park, IL), ADVIA Centaur (Siemens Healthcare Diagnostics, Deerfield, IL), Modular E170 (Roche Diagnostics, Indianapolis, IN), IMMULITE 2000 (Siemens Healthcare Diagnostics), and LIAISON (DiaSorin,...

متن کامل

Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.

OBJECTIVES To evaluate the newly developed Roche MODULAR Analytics E170 Total Vitamin D and the Siemens ADVIA Centaur Vitamin D Total assays. MATERIALS AND METHODS Assays were evaluated using the Clinical and Laboratory Standards Institute protocols. Split patient samples were compared with LC-MS/MS and DiaSorin LIAISON assays (n=79 including 15 specimens with detectable endogenous 25-OH vita...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2016