A simple method for evaluating stereotactic accuracy of Magnetic Resonance Imaging and Computed Tomography Imaging in Frame Based Radiosurgery.
نویسندگان
چکیده
PURPOSE To design a simple and reproducible method with minimum uncertainty for evaluating the stereotactic accuracy of Computed Tomography imaging and Magnetic Reso-nance Imaging and to compare the results. METHODS 3D phantom has been designed with 1 mm diameter targets in eight sectors for stereotactic MR imaging and CT imaging. The phantom was connected to Leksell stereotactic frame and the images obtained are exported to treatment planning system. Leksell stereotactic coordinates (X, Y, Z) of the holes are determined in each slice and mean maximum errors were calculated along with 3D vector distances from center of stereotactic coordinate system (100,100,100) to five known Targets. The target volumes are calculated independently for each image data sets. RESULTS The mean of maximum absolute error estimated for MR images from the Siemens Magnetom vision MR unit were 0.46 mm (X-Axis), 1.66 mm (Y-Axis) and 2.11 mm (Z-Axis). Mean of absolute maximum error estimated using CT images from the Philips Brilliance 16 CT scanner were 0.29mm (X-axis), 0.51mm (Y-Axis) and 0.90mm (Z-Axis). 3D vector calculations were 0.38 mm for CT and 0.72 mm for MRI. No significant variation is noticed in volume calculations from both image data sets. CONCLUSION This study showed that accuracy and quality of stereotactic CT- imaging and MR-imaging are limited by localization devices and their designs in frame based Radiosurgery. This method is simple, direct and economical with good resolution and high reproducibility. KEYWORDS Stereotactic accuracy, Computed Tomography, Magnetic Resonance Imaging, frame based Radiosurgery.
منابع مشابه
An image fusion study of the geometric accuracy of magnetic resonance imaging with the Leksell stereotactic localization system1
A special acrylic phantom designed for both magnetic resonance imaging (MRI) and computed tomography (CT) was used to assess the geometric accuracy of MRI-based stereotactic localization with the Leksell stereotactic head frame and localizer system. The acrylic phantom was constructed in the shape of a cube, 164 mm in each dimension, with three perpendicular arrays of solid acrylic rods, 5 mm i...
متن کاملValidation of accuracy in image co-registration with computed tomography and magnetic resonance imaging in Gamma Knife radiosurgery
The latest version of Leksell GammaPlan (LGP) is equipped with Digital Imaging and Communication in Medicine (DICOM) image-processing functions including image co-registration. Diagnostic magnetic resonance imaging (MRI) taken prior to Gamma Knife treatment is available for virtual treatment pre-planning. On the treatment day, actual dose planning is completed on stereotactic MRI or computed to...
متن کاملBimodal magnetic resonance imaging-computed tomography nanoprobes: A Review
Bimodal imaging combines two imaging modalities in order to benefit from their advantages and compensate the limitations of each modality. This technique could accurately detect diseases for diagnostic purposes. Nanoparticles simultaneously offer diagnostic data via various imaging modalities owing to their unique properties. Moreover, bimodal nanoprobes could be incorporated into theranostic s...
متن کاملGenerating Synthetic Computed Tomography and Synthetic Magnetic Resonance (sMR: sT1w/sT2w) Images of the Brain Using Atlas-Based Method
Introduction: Nowadays, magnetic resonance imaging (MRI) in combination with computed-tomography (CT) is increasingly being used in radiation therapy planning. MR and CT images are applied to determine the target volume and calculate dose distribution, respectively. Since the use of these two imaging modalities causes registration uncertainty and increases department w...
متن کاملDesign, Construction and Evaluation of an Anthropomorphic Head Phantom for Assessment of Image Distortion in Stereotactic Radiosurgery Planning Systems
Introduction: In recent years, the use of magnetic resonance (MR) images in radiation treatment planning has drawn considerable attention. However, although the extent of a tumor can be determined in great detail on MR images, the geometric accuracy of these images is limited by distortions stemming from the inhomogeneity of the static background magnetic field, the nonlineari...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Gulf journal of oncology
دوره 1 14 شماره
صفحات -
تاریخ انتشار 2013