Needle echogenicity in sonographically guided regional anesthesia: blinded comparison of 4 enhanced needles and validation of visual criteria for evaluation.

نویسندگان

  • Hans P Sviggum
  • Kyle Ahn
  • John A Dilger
  • Hugh M Smith
چکیده

OBJECTIVES Needle visualization is important for sonographically guided regional anesthesia procedures. Needle characteristics that improve needle visualization are therefore important to anesthesiologists. This study compared several echogenic needle designs by defining characteristics of needle echogenicity and assessing regional anesthesiologist preferences for these characteristics across various needle angles. METHODS Twelve blinded regional anesthesiologists graded 5 randomized block needles (1 nonechogenic control and 4 echogenic) on 4 predefined characteristics (overall brightness of the needle, overall clarity of the needle, brightness of the needle tip, and clarity of the needle tip). In-plane needle images in a gel phantom were obtained at 4 needle angles (15°, 30°, 45°, and 60°). Participants rated specific needle characteristics for each needle at each angle and then ranked their overall needle preferences. RESULTS Significant differences in all 4 needle characteristics were found across needle types (P< .01). Clarity of the needle tip was significantly associated with overall needle rank (P = .009). Other needle visualization characteristics were not significantly correlated with needle rank. The SonoPlex Stim needle (Pajunk Medical Systems, Tucker, GA) was rated highest in all 4 predefined needle characteristics as well as overall needle rank. CONCLUSIONS This study shows that anesthesiologists prefer certain visual characteristics of needles used in sonographically guided regional anesthesia procedures. Specifically, needle tip clarity most closely predicted clinician needle preferences. These results support the idea that all echogenic needle designs do not uniformly enhance needle visualization. Further studies are needed to determine whether needles with superior tip clarity predict not only clinician preferences but also improved sonographically guided regional anesthetic outcomes.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A review of the benefits and pitfalls of phantoms in ultrasound-guided regional anesthesia.

With the growth of ultrasound-guided regional anesthesia, so has the requirement for training tools to practice needle guidance skills and evaluate echogenic needles. Ethically, skills in ultrasound-guided needle placement should be gained in a phantom before performance of nerve blocks on patients in clinical practice. However, phantom technology is varied, and critical evaluation of the image...

متن کامل

Needle echogenicity in ultrasound-guided lumbar spine injections: a cadaveric study.

BACKGROUND Echogenicity of regional anesthesia needles has been tested on different preclinical models; however, previous studies were done in an ideal experimental setting utilizing high-frequency insonation and superficially located targets. Because steep-angle deep injections are typically required for spinal and other chronic pain procedures, and low-frequency transducers are used, further ...

متن کامل

Do echo-enhanced needles make a difference in sonographically guided vascular access?

OBJECTIVES The purpose of this study was to compare sonographically guided vascular access using standard and echo-enhanced needles in a variety of tissue-simulating vascular phantoms. METHODS We conducted a prospective single-blinded observational study at an academic medical center. All participants performed real-time sonographically guided vascular access using both a standard 18-gauge ne...

متن کامل

Comparison of Inferior Dental Nerve Block Injections in Child Patients Using 30-Gauge and 27-Gauge Short Needles

Introduction: The purpose of the present study was to record their pain sensation and to assess children`s reaction objectively and subjectively while receiving dental local anesthesia with 27- and 30-gauge needles. Methods: Forty children (24 boys and 16 girls) participated in this study. A random double-blinded crossover design was used so that each child served as his or her own control, rec...

متن کامل

Arterial blood gases from central venous lines: a sign for malformation.

Single-operator real-time ultrasound-guidance to aim and insert a lumbar epidural needle. curved needle set refines ultrasound guided ganglion impar neuro-lysis. Needle echogenicity in sono-graphically guided regional anesthesia: blinded comparison of 4 enhanced needles and validation of visual criteria for evaluation. 'Do as you would be done by': the ethics of using outdated equipment in medi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine

دوره 32 1  شماره 

صفحات  -

تاریخ انتشار 2013