A simulation study of the effect of drift electric fields on the response of radiation detectors using the PENELOPE code
نویسنده
چکیده
The effect of the presence of a drift electric field on the response of gaseous and semiconductor radiation detectors to energetic X-rays (energies Eph from 20 to 200 keV) is investigated using the PENELOPE code to simulate the photo-absorption and the slow-down of the electrons produced in Si, Ge, and Xe gas at 1 atm. For typical drift fields, the energy Ed deposited in the detection media is calculated taking into account the energy exchanged by the electrons with the field. The analysis of the calculated Ed distributions shows that the effect of the field on the distributions is negligible in Si and Ge semiconductor detectors, but not in Xe gas detectors, where for E=p 1⁄4 0:8V cm 1 Torr 1 the fluctuations introduced by the field for Eph 180 keV approach the intrinsic values for Xe, and the intrinsic discontinuity in linearity when Eph crosses the Xe K-edge (34.56 keV) is further reduced by 4%. The simulation data also suggest that this field effect may cause some deviations to the expected Gaussian response of Xe detectors to the absorption of monoenergetic photons. r 2006 Elsevier B.V. All rights reserved. PACS: 29.40. n; 29.40.Cs; 29.40.Wk; 29.30.Kv
منابع مشابه
Estimation of Dosimetric Parameters based on KNR and KNCSF Correction Factors for Small Field Radiation Therapy at 6 and 18 MV Linac Energies using Monte Carlo Simulation Methods
Background: Estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. Hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at differen...
متن کاملMonte Carlo Simulation of Prostate Intensity Modulated Radiotherapy Using PRIMO Software: A Feasibility Study
Introduction: Nowadays Intensity Modulated Radiotherapy (IMRT) is a common method for treating prostate cancers. Must of the Monte Carlo software cannot simulate the IMRT procedures due to inability of these soft wares to simulate the multi leaf collimator (MLC) positions or movements. A new user-friendly software based on the PENELOPE Monte Carlo code named PRIMO was published...
متن کاملThe plastic tray effect on the spectrum of photoneutrons prudused by 18 MV medical linac using the Monte Carlo method
Introduction: One of the important problems of using high energy linacs is the production of photoneutrons. High energy medical linacs produce secondary neutrons. These phtoneutrons increase patient dose Materials and Methods: In this study, the effect of plastic trays on spectrum of photoneutrons produced by 18MV linac is investigated using the MCNP4C code...
متن کاملPassive neutron area monitor with CR39
Background: In high-intensity, mixed and pulsed neutron fields the use of spectrometers or area monitors with active detectors is useless in these conditions neutron measuring devices must have a passive detector. Here a passive neutron area monitor with CR39 track detector was designed and the response was calculated. Materials and Methods: The response of a passive neutron area monitor with C...
متن کاملInvestigation of LINAC Structural Effects on Photoneutron Specified Parameters Using FLUKA code
Introduction: The utilization of high-energy photons in the medical linear accelerator can lead to photoneutron production. This study aimed to evaluate the effect of the physical components of the head, including flattening filter (FF) andmultileaf collimator (MLC), as well as the dependence of therapeutic field size on the photoneutron spectrum, dose, and flux. <str...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2006