Computed Radiography dose optimization in pediatric patients

نویسندگان

  • B. Juste
  • R. Tortosa
  • J. I. Villaescusa
  • G. Verdú
چکیده

Radiation dose reduction in pediatric X-ray imaging is especially important because of children radiation sensitivity. For any radiographic examination performed at a fixed radiographic tube potential, the patient absorbed dose is directly proportional to the value of milliampere-seconds (mAs) selected by the operator. Nevertheless, reducing X-ray exposure has the unavoidable disadvantage of increasing the quantum noise in the resultant image. The objective of this work is to identify the minimum tube current setting required for maintaining accurate examinations, to modify, if required, the daily protocols applied at La Fe de Valencia Universitary Hospital. To accomplish this goal, a noise addition software has been developed in order to study the diagnostic accuracy as a function of reducing dose by artificially increasing the image noise. The noise addition tool has been applied to several thorax images acquired from pediatric unit to simulate new lower dose radiographies and allow medical researchers to study how lower dose affects the patient pneumonia diagnosis.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Estimating and comparing the radiation cancer risk from cone-beam computed tomography and panoramic radiography in pediatric and adult patients

Background: This study aimed to estimate and compare the absorbed dose, lifetime cancer risk and mortalities due to cone-beam computed tomography (CBCT) and panoramic examinations on patients. Materials and Methods: The exposure factors were applied to 332 patients in two age groups (6-10, and ˃18-year-old). The dose-area product (DAP) values were measured for CBCT and panoramic radiographies. ...

متن کامل

Optimization of Dose and Image Quality in Full-fiand Computed Radiography Systems for Common Digital Radiographic Examinations

IntroductionA fine balance of image quality and radiation dose can be achieved by optimization to minimize stochastic and deterministic effects. This study aimed in ensuring that images of acceptable quality for common radiographic examinations in digital imaging were produced without causing harmful effects. Materials and MethodsThe study was conducted in three phases. The pre-optimization inv...

متن کامل

Evaluation of the Entrance Surface Dose (ESD) and Radiation Dose to the Radiosensitive Organs in Pediatric Pelvic Radiography

Background Patients' dosimetry is crucial in order to enhance radiation protection optimization and to deliver low radiation dose to the patients in a radiological procedure. The aim of this study was to assess the entrance surface dose (ESD) and radiation dose to the radiosensitive organs in pediatric pelvic radiography. Materials and Methods The studied population included 98 pediatric patien...

متن کامل

Efficacy of Increasing Focus to Film Distance (FFD) for Patient’s Dose and Image Quality in Pediatric Chest Radiography

Background: Increasing the x-ray focus to film distance (FFD) has been advocated as an effective method to reduce the patients’ radiation dose. The aim of this study was to investigate the efficacy of this technique in patients' dose and image quality in pediatric chest x-ray.Material and methods: Sixty pediatric patients were x-ray imaged at FFDs of 100 and 130 cm. Dose measurements were perfo...

متن کامل

Determination of Diagnostic Reference Level in routine Computed radiography examination in north of Iran

Introduction: Digital radiography has a major role in nowadays healthcare system so radiation protection (justification and optimization) is important in this field. Many related organizations to protection suggested to establish national and regional diagnostic reference level for medical procedures. Reference levels are typically set at the 75th percentile of the dose distri...

متن کامل

Optimization of Radiological Protection in Pediatric Patients Undergoing Common Conventional Radiological Procedures: Effectiveness of Increasing the Film to Focus Distance (FFD)

Background Increasing the x-ray film to focus distance (FFD), has been recommended as a practical dose optimization tool for patients undergoing conventional radiological procedures. In the previous study, we demonstrated a 32% reduction in absorbed dose is achievable due to increasing the FFD from 100 to 130 cm during pediatric chest radiography. The aim of this study was to examine whether in...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2008