Twiddler's syndrome complicating a transvenous defibrillator lead system.
نویسندگان
چکیده
Twiddler's syndrome is a rare complication seen in patients with implanted pacemakers or defibrillators. The condition typically presents with device malfunction and occurs when the patient either consciously or unconsciously twists and rotates the implanted device in its pocket, resulting in torsion and dislodgement of the implanted lead. A case of twiddler's syndrome involving a transvenous defibrillation lead and an abdominally implanted defibrillator is described. This is the first report of this complication with this particular lead. The patient in this report was a middle-aged obese diabetic woman who presented 7 months after defibrillator implantation with device noncapture and intermittent nonsensing. Review of the literature reveals that the majority of patients with this complication are middle-aged obese women with a defibrillator pocket that exceeds the size of the defibrillator. Treatment measures are discussed both for the patient with this complication and for the patient at increased risk for its occurrence.
منابع مشابه
"Twiddler" In A Canine Model And A Modified Implant Technique To Prevent It
This report explains the Twiddler's syndrome, a well known human entity, in a canine model. We also describe a modified technique to implant internal cardioverter defibrillators (ICD) in dogs to prevent Twiddler's syndrome. Our original study was to identify the factors influencing the DFT measurement and a canine model was selected as it is the best proven electrophysiological model. During th...
متن کاملLife-Threatening Effect of Twiddler's Syndrome with a Subpectorally Implanted Cardioverter Defibrillator
Twiddler's syndrome has been known for over 30 years as a rare complication after pacemaker implantation [1]. The often unintended manipulation of the pacemaker by patients causes the device to rotate in the pacemaker pocket, which may result in lead defects or dislocations. The same mechanism exists for transvenously implanted defibrillators and has been described as a rare complication for ab...
متن کاملTwiddler's syndrome in a patient with a single-chamber implantable cardioverter defibrillator.
A 67-year-old, obese, Caucasian female with a history of ischaemic dilated cardiomyopathy and ventricular fibrillation, underwent a single-chamber implantable cardioverter defibrillator (ICD) with a screw, in ventricular lead implantation. Appropriate sensing, pacing, and defibrillation thresholds were obtained at implantation. A chest X-ray after implantation demonstrated a proper lead positio...
متن کاملSubcutaneous defibrillator implantation in pediatric patients
OBJECTIVE Although sudden cardiac death is rare in children, an intracardiac defibrillator system is indicated in children with various types of cardiomyopathy, primary electrical diseases, and after surgical repair of congenital heart defects. The use of transvenous defibrillator lead systems is limited in pediatric patients because of a small body size and/or limited vascular access. Subcutan...
متن کاملA permanent transvenous lead system for an implantable pacemaker cardioverter-defibrillator. Nonthoracotomy approach to implantation.
A transvenous lead system for implantable defibrillators would obviate a surgical thoracotomy and reduce the morbidity and mortality associated with implantation. We evaluated the clinical performance of a new nonthoracotomy lead system that included a defibrillation lead in the coronary sinus. At the time of defibrillator implantation, transvenous defibrillation leads were inserted percutaneou...