Assessment of recently developed blood gas analysers: a multicentre evaluation
نویسندگان
چکیده
Providing guidelines for testing expected inaccuracy and imprecision is still a matter under debate. The Expert Panel of the French Society of Clinical Chemistry has developed a protocol, which was based on a comparative multi-centre evaluation of four instruments: the Ciba-Corning 278, the Instrumentation Laboratory 1306, the Nova SP 5 and the ABL 330. The purpose was to evaluate the analytical performance and efficiency of the analysers. Another aim was to design a valid approach for evaluating any new system. As buffered aqueous solutions and fluorocarbon emulsions give only partial information, tonometered blood was used at different levels of gas mixture, even though it is both difficult and time-consuming. Comparisons have been established on patients' blood samples with the analysers currently used in the evaluation sites. The tests showed that the four analysers have the same degree of precision, and interinstrument comparisons demonstrated a very high degree of reliability.This analysis emphasizes that the evaluation of instruments for pH and blood gas analysis is neither easy nor is it often done, mainly due to the choice of a quality-control material and the lability of the measured parameters.
منابع مشابه
Thermostatic regulation of blood samples in blood gas analysers: results and improved inethod applied to 11 different models
Thermostatic regulation of blood samples in blood gas analysers: results and improv ed inethod a ppli ed to 11 different models
متن کاملVariation of pH-measurement in platelet concentrates.
To measure pH in platelet concentrates, blood gas analysers with different calibration principles may be used. In this study, variances observed in pH measurements with two types of blood gas analysers were investigated. pH was measured in crystalloid solutions (platelet additive solution (PAS-II), phosphate-buffered solutions) and two types of platelet concentrates (containing 100% plasma, or ...
متن کاملAccuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion.
Although haematology analysers provide reliable full blood counts, they are known to be inaccurate at enumerating platelets in severe thrombocytopenia. If the thresholds for platelet transfusion, currently set at 10 x 10(9)/l, are to be further reduced, it is vital that the limitations of current analysers are fully understood. The aim of this large multicentre study was to determine the accura...
متن کاملAssessing the ABL 500 blood gas analyser
The ABL 500 blood gas analyser from Radiometer has cordless electrodes and does not use a humidifier for calibrating gases. During the evaluation of the analytical performance of this instrument, the problem of p0(2) accuracy was approached by comparing the values obtained with two kinds of tonometry (film and bubble). An acceptable level of imprecision was demonstrated for all measured paramet...
متن کاملInstrumental variability of respiratory blood gases among different blood gas analysers in different laboratories.
The aim of this study was to test the hypothesis that differences in oxygen tension (PO2) and carbon dioxide tension (PCO2) values from measurements performed on different blood gas analysers in different laboratories are clinically insignificant. Samples of fresh whole human tonometered blood (PO2 8.1 kPa (60.8 mmHg); PCO2 5.3 kPa (39.9 mmHg)) were placed in airtight glass syringes and transpo...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of Automatic Chemistry
دوره 11 شماره
صفحات -
تاریخ انتشار 1989