Does heparin coating improve patency or reduce infection of tunneled dialysis catheters?
نویسندگان
چکیده
BACKGROUND AND OBJECTIVES Tunneled dialysis catheters are prone to frequent malfunction and infection. Catheter thrombosis occurs despite prophylactic anticoagulant locks. Catheter thrombi may also serve as a nidus for catheter infection, thereby increasing the risk of bacteremia. Thus, heparin coating of catheters may reduce thrombosis and infection. This study evaluated whether heparin-coated hemodialysis catheters have fewer infections or greater cumulative survival than noncoated catheters. DESIGN, SETTING, PARTICIPANTS, & MEASUREMENTS We retrospectively queried a prospective access database to analyze the outcomes of 175 tunneled dialysis catheters placed in the internal jugular vein, including 89 heparin-coated catheters and 86 noncoated catheters. The primary outcome was cumulative catheter survival, and the secondary outcome was infection-free catheter survival. RESULTS The two patient groups were similar in demographics and clinical and catheter features. Catheter-related bacteremia occurred less frequently with heparin-coated catheters than with noncoated catheters (34 versus 60%, P < 0.001). Cumulative catheter survival was similar in heparin-coated and noncoated catheters (hazard ratio, 0.87; 95% confidence interval, 0.55 to 1.36; P = 0.53). On multiple variable survival analysis including catheter type, age, sex, diabetes, coronary artery disease, peripheral vascular disease, cerebrovascular disease, catheter location, and previous catheter, only catheter location predicted cumulative catheter survival (hazard ratio, 2.03; 95% CI, 1.27 to 3.25, with the right internal jugular location being the reference group, P = 0.003). The frequency of thrombolytic instillation was 1.8 per 1000 catheter-days in both groups. CONCLUSIONS Heparin coating decreases the frequency of catheter-related bacteremia but does not reduce the frequency of catheter malfunction.
منابع مشابه
Locking solutions for hemodialysis catheters; heparin and citrate--a position paper by ASDIN.
There is wide variation in the use of solutions to "lock" or fill tunneled central venous catheters for dialysis. Some centers use undiluted heparin concentrations ranging from 1000 to 10,000 U/ml and other centers place from 1000 to 10,000 U per lumen. Based on available evidence, it appears that heparin 1000 U/ml, or 4% sodium citrate are suitable choices for lock solution to maintain patency...
متن کاملLocking of tunneled hemodialysis catheters with gentamicin and heparin.
INTRODUCTION Catheter-related infection (CRI) is a major cause of morbidity and mortality in patients receiving hemodialysis. Antibiotic locking of these catheters has been shown to increase both the success of systemic antibiotic treatment in line sepsis, and to reduce the incidence of sepsis. We have studied the use of gentamicin locking of catheters (in combination with standard heparin rath...
متن کاملDoes the heparin lock concentration affect hemodialysis catheter patency?
BACKGROUND AND OBJECTIVES Concentrated heparin solutions are instilled into the catheter lumens after each hemodialysis session to prevent catheter thrombosis. The heparin lock concentration at many centers has been decreased recently to reduce the risk of systemic bleeding and contain costs. However, the effect of this change on catheter patency is unknown. We compared catheter patency between...
متن کاملSurvival of tunneled hemodialysis catheters after percutaneous placement.
BACKGROUND Tunneled catheters are becoming increasingly used as a permanent dialysis access. Easy way of insertion and good long-term patency make them competitive to fistulas in some groups of patients. METHODS Late complications and survival of 180 tunneled catheters inserted from June 2010 to December 2013 in 171 unselected hemodialysis patients were analyzed. RESULTS The cumulative time...
متن کاملA novel electrospun nano-fabric graft allows early cannulation access and reduces exposure to central venous catheters.
PURPOSE The use of tunneled central venous catheters (CVC) as vascular access for hemodialysis treatment is increasing worldwide. We present a novel polycarbonate urethane nano-fabric graft, produced by electrospinning technology, which has self-sealing features that avoid seroma formation and allow puncturing within 48 hours. The aim of this study was to assess its advantages in a setting wher...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical journal of the American Society of Nephrology : CJASN
دوره 4 11 شماره
صفحات -
تاریخ انتشار 2009