The Mechanical Effect of Rod Contouring on Rod-Screw System Strength in Spine Fixation
نویسندگان
چکیده
OBJECTIVE Rod-screw fixation systems are widely used for spinal instrumentation. Although many biomechanical studies on rod-screw systems have been carried out, but the effects of rod contouring on the construct strength is still not very well defined in the literature. This work examines the mechanical impact of straight, 20° kyphotic, and 20° lordotic rod contouring on rod-screw fixation systems, by forming a corpectomy model. METHODS The corpectomy groups were prepared using ultra-high molecular weight polyethylene samples. Non-destructive loads were applied during flexion/extension and torsion testing. Spine-loading conditions were simulated by load subjections of 100 N with a velocity of 5 mm min(-1), to ensure 8.4-Nm moment. For torsional loading, the corpectomy models were subjected to rotational displacement of 0.5° s(-1) to an end point of 5.0°, in a torsion testing machine. RESULTS Under both flexion and extension loading conditions the stiffness values for the lordotic rod-screw system were the highest. Under torsional loading conditions, the lordotic rod-screw system exhibited the highest torsional rigidity. CONCLUSION We concluded that the lordotic rod-screw system was the most rigid among the systems tested and the risk of rod and screw failure is much higher in the kyphotic rod-screw systems. Further biomechanical studies should be attempted to compare between different rod kyphotic angles to minimize the kyphotic rod failure rate and to offer a more stable and rigid rod-screw construct models for surgical application in the kyphotic vertebrae.
منابع مشابه
Effect of severity of rod contour on posterior rod failure in the setting of lumbar pedicle subtraction osteotomy (PSO): a biomechanical study.
BACKGROUND Rod failure has been reported clinically in pedicle subtraction osteotomy (PSO) to correct flat back deformity. OBJECTIVE To characterize the fatigue life of posterior screw-rod constructs in the setting of PSO as a function of the severity of rod contour angle. METHODS A modified ASTM F1717 to 04 was used. Rods were contoured to the appropriate angle for the equivalent 20-, 40-,...
متن کاملBiomechanical properties of anterior thoracolumbar multisegmental fixation: an analysis of construct stiffness and screw-rod strain.
STUDY DESIGN Three types of anterior thoracolumbar multisegmental fixation were biomechanically compared in construct stiffness and rod-screw strain. OBJECTIVES To investigate the effects of rod diameter and rod number on construct stiffness and rod-screw strain in anterior thoracolumbar multisegmental instrumentation. SUMMARY OF BACKGROUND DATA No studies have been undertaken to investigat...
متن کاملDesign and performance of spinal fixation pedicle screw system.
Pedicle screw-rod bilateral constructions are extensively used in spinal fixation. In this study, the common cause for failure of bilateral constructions has been determined to be the high stress concentration at the rod-setscrew interface. In order to overcome this problem, a design modification has been made by using a supplementary part (shoe) between rod and setscrew. Performance comparison...
متن کاملThe effects of rod contouring on spinal construct fatigue strength.
STUDY DESIGN In vitro fatigue loading using a corpectomy model outfitted with posterior pedicle screw instrumentation. OBJECTIVE The purpose of this study was to detect differences in fatigue resistance of titanium and stainless steel spinal constructs that use rods contoured using a French Bender, and to compare differences in fatigue resistance of contoured and straight titanium rods. SUM...
متن کاملEffect of Pedicle Fill on Axial Pullout Strength in Spinal Fixation After Rod Reduction.
Rod reduction to pedicle screws is used for a variety of spinal fixation procedures; however, it can alter the integrity of the screw-bone interface. The authors investigated the effect of pedicle fill (ratio of pedicle screw diameter to pedicle diameter) on the strength of the screw-bone interface after simulated rod reduction on 17 vertebrae (3 thoracolumbar spine specimens). Pedicle diameter...
متن کامل