Evaluation of Effective Sources in Uncertainty Measurements of Personal Dosimetry by a Harshaw TLD System
نویسندگان
چکیده
BACKGROUND The accurate results of the individual doses in personal dosimety which are reported by the service providers in personal dosimetry are very important. There are national / international criteria for acceptable dosimetry system performance. Objective : In this research, the sources of uncertainties are identified, measured and calculated in a personal dosimetry system by TLD. METHOD These sources are included; inhomogeneity of TLDs sensitivity, variability of TLD readings due to limited sensitivity and background, energy dependence, directional dependence, non-linearity of the response, fading, dependent on ambient temperature / humidity and calibration errors, which may affect on the dose responses. Some parameters which influence on the above sources of uncertainty are studied for Harshaw TLD-100 cards dosimeters as well as the hot gas Harshaw 6600 TLD reader system. RESULTS The individual uncertainties of each sources was measured less than 6.7% in 68% confidence level. The total uncertainty was calculated 17.5% with 95% confidence level. CONCLUSION The TLD-100 personal dosimeters as well as the Harshaw TLD-100 reader 6600 system show the total uncertainty value which is less than that of admissible value of 42% for personal dosimetry services.
منابع مشابه
Evaluation of Effective Sources in Uncertainty Measurements of Personal Dosimetry by a Harshaw TLD System
Background: The accurate results of the individual doses in personal dosimety which are reported by the service providers in personal dosimetry are very important. There are national / international criteria for acceptable dosimetry system performance.Objective: In this research, the sources of uncertainties are identified, measured and calculated in a personal dosimetry system by TLD.Method: T...
متن کاملTwenty Five Years of TL-Dose Measurements at the NRC-NEGEV
In the early years, the routine personal dosimetry at the Nuclear Research CenterNegev was performed by using photographic films. In 1975, the TLD method was introduced. The routine personal dosimetry was performed by using the LiF:Mg, Ti phosphors (G-1 cards), and employing an automatic Harshaw 2271 reader. The heating of the phosphor was performed by a hot finger, and the integral light outpu...
متن کاملبررسی مقایسهای پاسخدهی و رفتار دزسنجهای گرماتاب فردی نوترون-گاما در میدانهای 241Am-Be و 252Cf
Neutron individual dosimetry is challenging subject in dosimetry field. Thermoluminescent dosimeter (TLD) is one of the types of personal dosimeters that is extremely used to determine the doses of neutron sources. Typically, this type of dosimeter is used in different centers with a variety of applications after calibration with a standard neutron radiation sources. Usually, materials of TL cr...
متن کاملAchievement of Quality Control Parameters for NIS Personal Thermoluminescence Dosimeter (TLD) System
This work intend to identify and follow up quality control parameters to improve thermoluminescence dosemeter (TLD) system used for personal dosimetry in ionizing radiation metrology laboratory (IRML). Characterization of personal dosimetry system has presented in terms of the stability of reader calibration factors, linearity of response for Cs as reference radiation source, energy and angular...
متن کاملتولید دزیمتر LiF: Mg, Ti و مقایسه پاسخ آن با دزیمتر (LiF: Mg, Ti (TLD-100 ساخت شرکت هارشا در برابر پرتوهای گاما
Thermoluminescence dosimeters are small tablets used with 3. 2*3. 2*0. 9mm^3 for measurement of received dose from radioactive various beams. The most common dosimeter is TLD-100 which is lithium-fluoride family and this dosimeter contain magnesium impurities. In this study, first lithium-flouride powder was mixed with titanium and controlled atmosphere samples were heated. After measuring ...