Clinical evaluation of the pressure phosphene tonometer in patients with glaucoma.

نویسندگان

  • E Rietveld
  • D A van den Bremer
  • H J Völker-Dieben
چکیده

AIM To evaluate the reliability of the pressure phosphene tonometer in comparison with the Goldmann applanation tonometer. METHODS 45 consecutive patients with glaucoma (78 eyes) participated in the study. Eyes with previous eye surgery, a documented peripheral visual field defect, a refractive error of more than 5 dioptres, and patients who were unable to understand the procedure were excluded from the study. Intraocular pressure was measured with a pressure phosphene tonometer by one examiner and with a Goldmann applanation tonometer by two other examiners no more than 15 minutes apart. A second series of measurements was performed several weeks later on 34 patients (59 eyes). There was no communication between examiners or between examiner and patient regarding test results. RESULTS No statistically significant correlation was found between the applanation tonometry values and those obtained with a pressure phosphene tonometer. CONCLUSION The pressure phosphene tonometer is not suitable for reliably measuring intraocular pressure.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients.

PURPOSE To evaluate whether the pressure phosphene tonometer (PPT) is suitable for self-tonometry in patients with glaucoma or ocular hypertension. METHODS This was a prospective comparative study of 102 eyes of 102 patients with chronic glaucoma or ocular hypertension. Intraocular pressure (IOP) measurements by the Goldmann tonometer (GT) were compared with self-measured readings with the PP...

متن کامل

The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice.

PURPOSE To evaluate the Proview Eye Pressure Monitor as a medical instrument and as a technique for enabling a patient to obtain an accurate measure of his or her intraocular pressure (IOP). DESIGN An experimental laboratory evaluation and an independent prospective clinical study to test the reproducibility and accuracy of the Proview technique relative to Goldmann applanation tonometry. P...

متن کامل

Intraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea

Background: To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ. Methods: Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal chil...

متن کامل

Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

BACKGROUND To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. DESIGN A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES One hundred nine (54 males : 55 females) patients includ...

متن کامل

Perkins hand-held applanation tonometer. A clinical evaluation.

MANY ophthalmologists measure intra-ocular pressure with the Goldmann tonometer. It is neither portable nor capable of being used with the patient supine, which is disadvantageous in glaucoma screening programmes when it is often necessary to examine sick patients in their homes. An applanation tonometer, which is portable and can be used with the patient supine, sitting, or even standing, has ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The British journal of ophthalmology

دوره 89 5  شماره 

صفحات  -

تاریخ انتشار 2005