Evaluation of unintended electrical stimulation from MR gradient fields.

نویسندگان

  • Howard I Bassen
  • Leonardo M Angelone
چکیده

Exposure of patients with active implants (e.g. cardiac pacemakers and neurostimulators) to magnetic gradient fields (kHz range) during magnetic resonance imaging presents safety issues, such as unintended stimulation. Magnetically induced electric fields generate currents along the implant's lead, especially high at the distal tip. Experimental evaluation of the induced electric field was previously conducted. This study aimed to perform the same evaluation by means of computational methods, using two commercially available software packages (SemcadX and COMSOL Multiphysics). Electric field values were analyzed 1-3 mm from the distal tip. The effect of the two-electrode experimental probe was evaluated. The results were compared with previously published experimental data with reasonable agreement at locations more than 2-3 mm from the distal tip of the lead. The results were affected by the computational mesh size, with up to one order of magnitude difference for SEMCAD (resolution of 0.1 mm) compared to COMSOL (resolution of 0.5 mm). The results were also affected by the dimensions of the two-electrode probe, suggesting careful selection of the probe dimensions during experimental studies.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In-vitro mapping of E-fields induced near pacemaker leads by simulated MR gradient fields

BACKGROUND Magnetic resonance imaging (MRI) of patients with implanted cardiac pacemakers is generally contraindicated but some clinicians condone scanning certain patients. We assessed the risk of inducing unintended cardiac stimulation by measuring electric fields (E) induced near lead tips by a simulated MRI gradient system. The objectives of this study are to map magnetically induced E near...

متن کامل

Analytic Calculation of the E-Fields induced by Gradient Coil Switching

Introduction: Nerve stimulation induced by rapidly switching magnetic field gradients poses an upper limit on the gradient strengths that can be employed in fast magnetic resonance imaging and diffusion measurements’. The likelihood of stimulation is usually assessed by evaluation of the maximum rate of change of the magnetic field, inside the gradient coi12. Nerve stimulation actually occurs a...

متن کامل

Effect of phasic electrical locus coeruleus stimulation on inhibitory and excitatory receptive fields of layer V barrel cortex neurons in male rat

Introduction: It is believed that Locus Coeruleus (LC) influences the sensory information processing. However, its role in cortical surround inhibitory mechanism is not understood. In this experiment, using controlled mechanical displacement of whiskers we investigated the effect of phasic electrical stimulation of LC on response of layer V barrel cortical neurons in anesthetized rat. Methods: ...

متن کامل

Magnetic resonance acoustic radiation force (impulse) imaging (MR-ARFI)

Background/introduction MR-ARFI adds to the rich toolbox of MR imaging methods for guiding focused ultrasound. MRI has exquisite soft tissue contrast for targeting focused ultrasound and a variety of contrast mechanisms for assessing the effect of focused ultrasound. In addition, MR can provide temperature mapping in near-realtime. MR-ARFI is an imaging method that compliments these other capab...

متن کامل

Consideration of magnetically-induced and conservative electric fields within a loaded gradient coil.

We present a method to calculate the electric (E)-fields within and surrounding a human body in a gradient coil, including E-fields induced by the changing magnetic fields and "conservative" E-fields originating with the scalar electrical potential in the coil windings. In agreement with previous numerical calculations, it is shown that magnetically-induced E-fields within the human body show n...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Frontiers in bioscience

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2012