Radon monitor calibration using NIST radon emanation standards: steady flow method.
نویسندگان
چکیده
The National Institute of Standards and Technology (NIST) polyethylene-encapsulated 226Ra/222Rn emanation (PERE) standards (old SRM 4968 and new SRMs 4971, 4972, and 4973) provide precise radon emanation rate, certified to a high degree of accuracy (approximately to 2%). Two new SRM 4973 standards containing totally 1036 Bq (0.028 microCi) of 226Ra, emanate 0.114 Bq (3.08 pCi) of 222Rn per min. Air passing over such sources at a flow rate of 1 l min(-1) will have a radon concentration of 114 Bq m(-3) (3.08 pCi l(-1)). This paper describes a practical calibration system and the actual calibration verification data obtained at different flow rates, for E-PERM passive radon monitors, Femto-Tech and Alpha Guard Continuous Radon Monitors. The use of such an affordable and easy to use system by the manufacturers and users of radon measurement devices will bring uniform standards with traceability to a NIST standard source and is considered an important step in standardising radon measurement methods.
منابع مشابه
Nist Traceable Radon Calibration System for Calibrating
The NIST (National Institute of Standards and Technology) has recently made available 222Rn emanation standards for test and evaluation. The NIST certified parameters include the ^ ~ a strength and the emanation coefficient. When such a source is loaded into a radon leak tight enclosure of a known volume, 222~n accumulates over time and it is possible to calculate precisely the time integrated ...
متن کاملAPPLICATION OF NIST ^Rn EMANATION STANDARDS FOR CALIBRATING ^Rn MONITORS
Several years of research in developing a radon emanation standard by the National Institute of Standards and Technology (NIST) has led to NIST Radon Emanation Standard SRM 4968 (Journal of Research of the MST 100:629-639; 1996). This new standard is based on a polyethylene -encapsulated ""Ra solution that has been demonstrated to emanate a well-characterized and known quantity "̂Rn when employe...
متن کاملRadon diffusion and the emanation fraction for NIST polyethylene capsules containing radium solution
The national standard for radon (Rn) measurements in the United States is based on Standard Reference Materials (SRMs) prepared at the National Institute of Standards and Technology (NIST), Gaithersburg, MD. NIST radon standards are hermetically sealed polyethylene capsules, filled with radium (Ra) solution. In preparation of standard reference materials for Rn, it is necessary to understand pr...
متن کاملCalibration of Electret-Based Integral Radon Monitors Using NIST Polyethylene-Encapsulated 226Ra/222Rn Emanation (PERE) Standards
The recently developed 222Rn emanation standards that are based on polyethylene-encapsulated 226Ra solutions were employed for a first field-measurement application test to demonstrate their efficacy in calibrating passive integral radon monitors. The performance of the capsules was evaluated with respect to the calibration needs of electret ionization chambers (E-PERM®, Rad Elec Inc.). The enc...
متن کاملThe NIST Primary Radon-222 Measurement System
Within the United States, the national standard for radon measurements is embodied in a primary radon measurement system that has been maintained for over 50 years to accurately measure radon (222Rn) against international and national radium (226Ra) standards. In turn, all of the radon measurements made at the National Institute of Standards and Technology (NIST) and the radon transfer calibrat...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Radiation protection dosimetry
دوره 113 1 شماره
صفحات -
تاریخ انتشار 2005