Shock-induced right ventricular pacing failure caused by a short circuit: Uncommon but life-threatening complication of the Riata lead
نویسندگان
چکیده
Introduction The Riata and Riata ST implantable cardioverter-defibrillator (ICD) leads, a narrow caliber silicone lead without an outer jacket, have been shown to be prone to structural and electrical failure. While much interest has been shown in potentially lethal high-voltage (HV) failures caused by electrical short circuits due to a breach in the ethylene tetrafluoroethylene (ETFE) insulation of the defibrillation cables, the clinical manifestation of Riata/ST lead failures has not been fully elucidated. In this report, we present a case of shock-induced Riata lead pacing failure and discuss clinical approach to the management of Riata/ST leads.
منابع مشابه
Externalized Conductor Cables in QuickSite Left Ventricular Pacing Lead and Riata Right Ventricular Lead in a Single Patient: A Common Problem With Silicone Insulation
QuickSite (St Jude Medical, Sylmar, CA, USA) is a silicone and polyurethane-insulated coronary sinus pacing lead. Riata lead (St Jude Medical, Sylmar, CA, USA) is a silicone insulated right ventricular shock lead. Recently, insulation breach of silicone based leads raised a huge concern. Fluoroscopic examination of these two leads in the same patient revealed externalization of these two leads....
متن کاملRiata silicone defibrillation lead with normal electrical measures at routine ambulatory check: The role of high-voltage shock testing
AIM To describe our experience with shock testing for the evaluation of patients with Riata™ leads. METHODS Among 51 patients with normal baseline electrical parameters, 20 died during follow-up. Of the remaining 31 patients, 15 underwent the test: In 10 cases a defibrillation testing with ventricular fibrillation (VF) induction and in 5 cases a R-wave-synchronized shock (> 20 J, without indu...
متن کاملApplication of centrifugal pump in right ventricular failure due to pulmonary artery hypertension. (A case report)
Background: It has not been yet developed a ost suitable method to treat right ventricular failure due to pulmonar artery hypertention resulted from ventricular arrhythia. Though some case reports about patients ready to heart transplant using left assist device or biventral assist device as bridge have been published in Journals, but nocase concerning application of centrifugal pump for treatm...
متن کاملDiathermy-induced ventricular fibrillation with Riata high-voltage lead insulation failure.
A 76-year-old man presented after a series of six shocks in quick succession from his implanted biventricular defibrillator, without relief of palpitations and pre-syncope. Initial 12-lead electrocardiogram revealed ventricular tachycardia at 180 b.p.m. One 200 J external shock was delivered with successful return to sinus rhythm. Interrogation of the device revealed that the device had been re...
متن کاملUnusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy.
We describe the case of a patient who presented with multiple implantable cardioverter-defibrillator (ICD) shock discharges 12 months after device implantation. Upon device interrogation, intermittent oversensing of electrical noise and potential ICD lead failure were suggested by a significant increase in the sensing integrity counter (SIC), a cumulative count of very short ventricular sensed ...